Page 8
2. Operating Modes
The function switch is used to select the required mode. The four operating modes
available are…
1.
SET LIMIT
: Switch to
SET LIMIT
when setting the Redox value at which
the Alarm LED and (optional) relay output are activated. The
relay output is not operational in this mode. See section 3.5.
2.
STANDBY
: Switch to
STANDBY
to inhibit the Alarm LED and (optional)
output relay action. Select this mode for calibration, or when
automatic dosing is not required.
3.
AUTO
: Switch to
AUTO
to enable the Alarm LED and (optional)
relay output. These will be activated when the Redox is
outside the limit value. See section 3.5.
4.
MANUAL
: Switch to
MANUAL
to manually dose chemicals when the
optional relay output is fitted. The relay output will be
activated for as long as the
miniCHEM-Redox
is in this
mode, regardless of the Redox reading. See section 7.