Display 

Custom Liquid Crystal Display, shows Relative Humidity & 
Temperature simultaneously. Indicators show the current mode. 

-

 

On / Off Switch (

). 

Press to switch the 

MC-83

 on. Press a second time to switch the unit 

back off. Alternatively, the 

MC-83

 will switch itself off if no button has 

been pressed for 10 minutes. 

®

 

HOLD

 

Press to freeze a reading on the display. “

D.H

” is shown on the 

display whilst in Data Hold mode. Press a second time to un-freeze 
the display. This feature is useful in rapidly changing environments 
(see section 3).  

¯

 

o

C / 

o

F

 

Selects either 

O

Celsius or 

O

Fahrenheit units in Temperature readout 

mode. The display shows “

O

C” or “

O

F” accordingly. See section 2. 

°

 

RECORD

 

Press to start recording Maximum, Minimum and Average values. 
Press a second time to stop recording these values. See section 4 for 
a full description of this function. 

±

 

RECALL

 

Press to sequentially recall Maximum, Minimum and Average values 
recorded using the Record function. See section 4 for a full 
description of this function. 

²

 

Relative Humidity & Temperature Probe

 

The Relative Humidity & Temperature Sensor is conveniently 
supplied with 1 metre of cable. The humidity sensor is a high 
precision thin-film capacitance type for fast response. The thermistor 
type temperature sensor is also designed for fast response in air. 

³

 

Sensor INPUT socket.

 

Plug the sensor into this socket. Take care to align the plug correctly 
with the socket. 

´

 

Battery Compartment / Cover

 

Undo the retaining screw and slide out the cover to access the 
battery. 

Содержание MC-83

Страница 1: ...1 Table of Contents Each major section of the handbook is clearly listed Sub sections have also been included to enable you to find the information you need at a glance 2 Introduction The introduction has a diagram and explanation of the display and controls of the MC 83 It also contains a full listing of all of the items that you should have received with your MC 83 Please take the time to read t...

Страница 2: ...ty Ltd ABN 30 009 773 371 4 Jamberoo Street Springwood Brisbane Australia 4127 Phone 07 32 900 400 International 61 7 32 900 400 Fax 07 3808 4871 International 61 7 3808 4871 Email tps tps com au Web www tps com au ...

Страница 3: ... 6 1 3 Specifications 6 2 Relative Humidity and Temperature Measurement 7 3 Freezing the Reading on the Display 7 4 Data Record and Recall Functions 8 5 Battery Replacement 9 6 Display Mode Indicators 9 7 Humidity Calibration 10 7 1 Calibration Procedure 11 8 Troubleshooting 14 9 Warranty 15 ...

Страница 4: ...Page 4 1 Introduction 1 1 MC 83 Display and Controls ...

Страница 5: ...ingly See section 2 RECORD Press to start recording Maximum Minimum and Average values Press a second time to stop recording these values See section 4 for a full description of this function RECALL Press to sequentially recall Maximum Minimum and Average values recorded using the Record function See section 4 for a full description of this function Relative Humidity Temperature Probe The Relative...

Страница 6: ...andard 100mL 130091 1 3 Specifications Range Resolution Accuracy Humidity 10 to 95 RH 0 1 RH 3 RH Temperature O C 0 to 50 O C 0 1 O C 0 8 O C Temperature O F 32 to 122 O F 0 1 O F 1 5 O F Display Custom Liquid Crystal Display shows Relative Humidity and Temperature simultaneously Sensor Type Humidity Thin film capacitance sensor Temperature Thermistor Power 9V Alkaline Battery for 60 hours operati...

Страница 7: ...g in the bottom half of the display Note the O C or O F indicator on the display Press the O C O F button to change between O C or O F readout 3 Freezing the Reading on the Display When taking measurements in a rapidly changing environment or in a difficult to reach areas the operator is able to freeze the display This allows readings to be correctly noted 1 To freeze the display press the HOLD bu...

Страница 8: ...data record mode press the RECALL button The Maximum values will now be displayed Note the Max indicator on the display Press the RECALL button a second time The Minimum values will now be displayed Note the Min indicator on the display Press the RECALL button a third time The Average values will now be displayed Note the AVG indicator on the display Press the RECALL button a fourth time to return...

Страница 9: ...elow is a summary of the mode indicators that appear on the MC 83 display Indicator Meaning RH Indicates that the top reading is Relative Humidity O C Indicates that the bottom reading is Temperature in O C O F Indicates that the bottom reading is Temperature in O F D H Indicates that the MC 83 is in Data Hold mode see section 3 REC Indicates that the Record Data mode is currently active see secti...

Страница 10: ...ble 1 shows the Relative Humidity characteristics of these two standards When making up your own standards The solution must be saturated with the salt i e no further solids will dissolve Use distilled water Use a plastic bottle that allows the sensor tip to fit snugly into the top of the bottle making a good seal Allow 24 hours after mixing for the temperature to stabilise Standards can be stored...

Страница 11: ...art no 130091 The sensor head must be above the level of the standard as per the diagram below DO NOT IMMERSE THE SENSOR Leave for 3 hours to allow the Relative Humidity in the bottle to stabilise 2 Switch the meter on Wait for the displayed Relative Humidity reading to stabilise up to 3 minutes ...

Страница 12: ...hows the appropriate Relative Humidity value for the current temperature as detailed in Table 1 4 Fit the sensor head into a bottle of 75 humidity standard part no 130090 The sensor head must be above the level of the as per the diagram in Step 1 DO NOT IMMERSE THE SENSOR Leave for 3 hours to allow the Relative Humidity in the bottle to stabilise ...

Страница 13: ...ature as per Table 1 For example If A 70 0 Y 11 3 assuming the current temperature is 25 O C Z 75 3 assuming the current temperature is 25 O C Then B 70 0 x 75 3 11 3 70 0 11 3 B 70 0 x 64 0 58 7 B 76 3 6 Adjust VR4 see diagram in Step 3 above until the display shows the value of B calculated above 7 Adjust VR3 see diagram in Step 3 above until the display shows the value of Z above 8 Replace the ...

Страница 14: ...s Remedy Unstable readings 1 Battery is low 2 Requires re calibration 3 Sensor is faulty Replace battery Re calibrate meter Return to factory for repair Display shows just 1 Meter has over ranged 2 Instrument or sensor is faulty Check that the R H or temperature are within the specified ranges section 1 3 Return to factory for repair Meter will not turn on Battery is exhausted Replace the battery ...

Страница 15: ...the equipment Shipping damage is not covered by this warranty Please note A guarantee card is packed with the instrument or sensor This card must be completed at the time of purchase and the registration section returned to TPS Pty Ltd within 7 days No claims will be recognised without the original guarantee card or other proof of purchase This warranty becomes invalid if modifications or repairs ...

Страница 16: ...ime phone number Your company name ORDER number and return street address A description of the fault Please be specific Note Please Repair does NOT describe a fault Your equipment will be repaired and returned to you by air express where possible For out of warranty units a repair cost will be calculated from parts and labour costs If payment is not received for the additional charges within 30 da...

Отзывы: