Conductivity and Salinity Troubleshooting, continued...
Inaccurate readings,
even when
calibration is
successful.
1.
Electrode may have a
build-up of dirt or oily
material on electrode wires.
2.
Platinum-black coating has
worn off.
Clean electrode, as per the
instructions detailed in section
12.4.
Electrode requires
replatinisation.
Return to the factory, or see
details in section 12.5.
Readings drift.
1.
Electrode may have a
build-up of dirt or oily
material on electrode wires.
Clean electrode, as per the
instructions detailed in section
12.4.
Readings are low or
near zero.
1.
Electrode may have a
build-up of dirt or oily
material on electrode wires.
2.
Electrode is not immersed
deeply enough.
3.
Electrode is faulty.
Clean electrode, as per the
instructions detailed in section
12.4.
Immerse electrode at least to
the vent hole in the white
plastic cover.
Return electrode to factory for
repair or replacement.