10. Troubleshooting
10.1 General Error Messages
Error Message
Possible Causes
Remedy
Factory
Cal. Fail
See
Handbook
The EEPROM chip which
contains the factory calibration
information has failed.
The unit must be returned to
TPS for service.
•
Conductivity and Salinity
readings will be accurate after
a full two point calibration.
•
pH readings will be accurate
after a 2-point calibration (use
manual temperature
compensation).
•
Temperature readings may be
up to ±5
o
C incorrect.
Memory
Failed
Calib.
Lost
Memory
Reset !
You MUST
Re-Cal.
User calibration settings have
been lost or corrupted.
Re-calibrate the instrument.
Both a Zero and a Standard
calibration will be required for
Conductivity and Salinity (see
sections 3.1 and 4.1). A full 2
point calibration will be
required for pH (see section
5.1) and a 1 point calibration
for temperature (see section
6.1).
Meter displays the
word
OFF
, and
switches off.
Battery is below 6.00 volts.
Replace the battery.
Meter will not turn
on.
Battery is exhausted.
Replace the battery.
Flashing
symbol.
Battery is below 7.50 volts.
Replace the battery soon. Note
that the unit will switch itself
off when the battery falls below
6.00 volts.