9. Temperature Mode
The temperature readout must be calibrated before attempting pH or ppM Dissolved Oxygen
calibration and measurements.
The decimal point is replaced by a “
∗∗∗∗
” if the reading is not calibrated.
The
90-FLMV
is able to take Temperature readings from the Conductivity sensor or a separate
Temperature sensor. Only one or the other can be connected at any one time. Ensure that the
90-
FLMV
has been calibrated on the correct sensor.
9.1 Temperature Calibration
1.
Plug the Conductivity sensor or separate Temperature sensor into the
Cond/Sal
socket.
2.
Switch the meter on.
3.
Place the sensor into a beaker of room temperature water, alongside a good quality mercury
thermometer. Stir the sensor and the thermometer gently to ensure an even temperature
throughout the beaker.
4.
Select Temperature Calibration (
→
→
→
→
F1:Calibrate
→
→
→
→
F4:Temperature
).
The Temperature Calibration screen is now displayed…
Enter Actual Temperature : _ 24.0
O
c
Temperature Calibration Menu Quits
5.
The current reading from the sensor is displayed on the far right of the top line.
When this reading has stabilised, use the Numeric Keypad to enter the same temperature as
measured by the mercury thermometer.
6.
Press the
key to calibrate the temperature readout.
Alternatively, press the
key to abort temperature calibration.
The
90-FLMV
temperature is now calibrated. The following screen is displayed…
Calibration OK, Offset=0.1
o
c
The
90-FLMV
will now return to normal measurement mode and the “
∗∗∗∗
” in the Temperature
readout will have been replaced by a decimal point.