5. TDS Mode
5.1 Selecting TDS Mode
1.
Select TDS Mode (
→
→
→
→
F2:Mode
→
→
→
→
F3:TDS
).
2.
You will now the asked to enter the TDS factor. Type in the required factor between 0.40 and
1.00 and press
to save the value or
to quit.
The TDS factor will be reset to 0.65 when the meter is initialized.
Input TDS Factor: 0.65
3.
The
90-FLMV
is now in TDS measurement mode.
5.2 Setting the Conductivity sensor k factor
The
90-FLMV
automatically recognises a k=10 sensor.
The
90-FLMV
does not
automatically recognise k=0.1 or k=1 sensors. When a k=0.1 or k=1 sensor
is used, the
90-FLMV
must be set to the correct k factor before use.
To select a k=0.1 or k=1 sensor…
3.
Select k factor entry (
→
→
→
→
F4:Setup
→
→
→
→
F4:k factor
).
4.
The k factor entry screen is now displayed. The arrow indicates the current selection.
Select nominal k factor,
F1:k=.1 >F2:k=1
Press
if a k=0.1 sensor is being used.
Press
if a k=1 sensor is being used.
Press
to quit without changing the current setting.
Notes
4.
The manual k factor selection is reset to k=1 during initialisation.
5.
The
90-FLMV
will always automatically recognise a k=10 sensor, regardless of the manual k
factor selection.
6.
Calibration settings for k=0.1, k=1 and k=10 sensors are NOT stored separately.