16.5 Temperature Troubleshooting
Symptom
Possible Causes
Remedy
Temperature inaccurate and
cannot be calibrated.
1.
Faulty connector.
2.
Faulty Conductivity or
Temperature sensor,
whichever is being used.
Check the connector and replace if
necessary.
Return Conductivity or Temperature sensor
for repair, or replace sensor.
Displays manual
temperature setting (eg.
25.0
o
cM
)when
Conductivity or
Temperature sensor plugged
in.
1.
Faulty instrument socket.
2.
Faulty Conductivity or
Temperature sensor,
whichever is being used
Return the instrument to the TPS factory for
service.
Return Conductivity or Temperature sensor
for repair, or replace sensor.
NOTE
Conductivity and TDS readings may still be
accurate, as there is a separate temperature
sensor to provide ATC for these parameters
built into the Conductivity/TDS sensor.