background image

CALIBRATION

Turn on the PGM100 until the working display is illuminated. Wait for 2 minutes prior to
beginning the calibration process.

1. Prepare the approved gas cylinder and regulator with the calibration adapter attached.

The flow should be 200-400 cc/min.

2. Push and hold the Zero button until the display reads "GAS".

3. Attach the PGM100 calibration adapter. 

4. Start the gas flow. 

5. Push and release the Zero button. A changing number will flash as will the CAL icon. 

6. Successful calibration is indicated when only the gas concentration is displayed. The

CAL icon will no longer be illuminated. 

7. Bad calibration will be indicated by the word "BAD" and the CAL icon on the display.

Press the power button to acknowledge. (Retry calibration)

8. Successful calibration will be logged as an Alarm event in the log. 

BUMP/CALIBRATION using SCal-100 

(Sold Separately)

1. Prepare SCal-100 Station.

2. Turn PGM100 On and allow to Zero.

3. Install PGM100 into SCal-100 Station.

4. Remove PGM100 when test is complete

8

Содержание PGM100

Страница 1: ...PGM100 Carbon Monoxide Single Gas Monitor The Value LeaderTM www tpi thevalueleader com ...

Страница 2: ...Information 2 General Description Physical Specifications 3 Product Features 4 Sensor Specifications Alarm Settings Bump Testing 5 Adjust Date Time 6 Alarm Adjustment 7 Calibration 8 Operation and Use 9 10 Warranty and Repair Policy 11 ...

Страница 3: ...g Warning To reduce the risk of ignition of a flammable atmosphere bat teries must only be changed in an area known to be nonflammable Avertissement Pour reduire le risque d fallumage d fune atmosphere inflammable des batteries doivent seulement etre changees dans un secteur connu pour etre inflammables IP65 EMC Directive 2004 108 EC EN 61000 4 2 1995 EN 61000 4 3 2002 EN 55011 2007 Class 1 Groups...

Страница 4: ... environ ments above 21 oxygen 7 WARNING To verify operation prior to each day s use a function test bump test should be performed If the instrument does not pass this test full calibration should be performed 8 WARNING Keep all openings free from dirt debris and foreign objects 9 WARNING Do not use a damaged or improperly operating instru ment Contact a service representative immediately ONLY zer...

Страница 5: ...Figure 1 FUNCTION INDICATORS Battery Low Operation Time Remaining months Calibration High Low Alarm Set Points Factory preset to OSHA requirements TWA STEL Alarms Preset to OSHA limits Other alarm settings available Data logging Up to 100 alarm events Continuous ON Option PHYSICAL SPECIFICATIONS Operational Temp 4 to 122 F 20 to 50 C Humidity 15 90 RH Non condensing Duty Cycle Continuous Response ...

Страница 6: ...on Multiple LED s beneath the frosted area create a very bright warning even in bright sunlight The instrument is mounted using a sturdy suspender clip on the back of the instrument This instrument cannot be powered off or zeroed when in alarm mode All sensors and batteries are designed for 2 years of continuous normal use The on board memory will store up to 100 events or alarm condi tions These ...

Страница 7: ... and needs Fig 2 shows examples of some of the alarm ranges Fig 2 STANDARD ALARM SETTINGS Type Low High TWA STEL CO 50 200 50 75 See page 7 for Alarm Adjustment instructions FUNCTION BUMP TEST MANUAL Turn the instrument on and wait for the working display to illuminate Apply an approved concentration of gas to the sensor area using the PGM100 calibration adapter The flow should be 200 400 cc min T...

Страница 8: ...reased by pressing the right button and will roll over to 01 if it exceeds 12 Adjust to desired MO 7 Press and release the left button to accept it 8 PGM100 shows dd until either button is pressed and released 9 PGM100 shows the current date 10 The number is increased by pressing the right button and will roll over to 01 if it exceeds the maximum number of days for the month specified Adjust to de...

Страница 9: ...300 it will start from 0 6 Press Left button to save the alarm setting and it will now display HI 7 Press the Left button to adjust the HI alarm The current setting will be dis played 8 Press Right button to adjust the alarm You can only increase the number Once it exceeds 300 it will start from 0 9 Press Left button to save the alarm settings It will now display tA for TWA 10 Press Left button to...

Страница 10: ...ase the Zero button A changing number will flash as will the CAL icon 6 Successful calibration is indicated when only the gas concentration is displayed The CAL icon will no longer be illuminated 7 Bad calibration will be indicated by the word BAD and the CAL icon on the display Press the power button to acknowledge Retry calibration 8 Successful calibration will be logged as an Alarm event in the...

Страница 11: ... not gas free BAD will illuminate after the automatic zeroing followed by a reading based on factory zero set points Attach the instrument to the outermost garment as close to your head face as practical Follow federal state local and company regulations as it relates to the use of this prod uct Alarms will indicate unsafe levels The display will show concentration data in combina tion with audibl...

Страница 12: ...trument should be immediately taken out of service for battery replacement Only factory trained personnel should perform any internal servicing MO indicates months days remaining of operation time before battery and sensor servicing are required Only factory trained personnel should per form any internal servicing I2C indicates a communication error in the EEPROM Service is required Only factory t...

Страница 13: ...l use and does not cover damage which occurs in shipment or failure which results from alteration tampering accident mis use abuse neglect or improper maintenance Proof of purchase may be required before warranty is rendered Units out of warranty will be repaired for a service charge Internal repair of maintenance must be completed by a Test Products International autho rized technician Violation ...

Страница 14: ...Notes ...

Страница 15: ...Notes ...

Страница 16: ...intl com Test Products International Ltd 342 Bronte Road South Unit 9 Milton Ontario Canada L9T5B7 Tel 905 693 8558 Fax 905 693 0888 www testproductsintl com Test Products International Europe Ltd Longley House International Drive Southgate Crawley West Sussex RH10 6AQ Tel 44 0 1293 530196 Fax 44 0 1293 531870 www tpieurope com ...

Отзывы: