25
f. Continuity Buzzer
WARNING!
Do not attempt to make continuity measurements with
circuit energized.
Instrument set-up:
FUNC.
BLACK RED
MIN
MAX
TEST LEAD TEST LEAD
READING
READING
Ω
COM
V/
Ω
0.001K
Ω
4.000K
Ω
Measurement Procedure:
1.
Disconnect power to circuit to be measured.
2.
Plug black test lead into the
COM
input jack.
3.
Plug red test lead into
V/
Ω
input jack.
4.
Set the rotary switch to the
Ω
function.
5.
Press the
FUNC
button until is on the LCD.
6.
Connect test leads to circuit to be measured.
7.
Listen for the buzzer to confirm continuity.
Optional Modes
•
HOLD:
Freezes the reading on the LCD.
•
TRIM:
Push
TRIM
button to stabalize reading.
!
10
3. Specifications (cont.)
a. DCV
Range
Resolution
Accuracy
Impedance
40V
0.01V
±0.75% of reading,
10M
Ω
400V
0.1V
±3 digits
600V
1V
b1. ACV (45Hz to 60Hz)
Range
Resolution
Accuracy
Impedance
40V
0.01V
±1.0% of reading,
10M
Ω
400V
0.1V
±3 digits
600V
1V
b2. ACV
(20Hz to 45Hz, 60Hz to 1KHz)
Range
Resolution
Accuracy
Impedance
40V
0.01V
±2.5% of reading,
10M
Ω
400V
0.1V
±3 digits
600V
1V
*Warning:
Test Leads. Use only correct type and
overvoltage category rating.
!