background image

4. Basic Analyzer Controls 

4.1 Turning The Analyzer On

Always: 

- Before turning on please ensure that the Mini Pump Protection Filter

assembly and the Temperature Sampling Probe complete with In-Line Filter or the
Tubing & In-Line Filter for Leak Detection are 

not

connected to the Gas Sample

Port (see 2.2 or 2.4) 

Press and hold down the Power Key and the TPI 709R will start its 30 second
countdown 'PURGE' will be displayed: - The instrument MUST be turned on in a
clean air environment as the 30 second purge will set the Carbon Monoxide level
to Zero and the Oxygen to 20.9%. 
Ensure that the filters are clean and dry as dirty or wet filters will result in a loss
of flow rate and 'Lo Flo' will be displayed to inform you that filters should be
changed

During the last 20 seconds of the 30 second purge time the user can scroll
through the following Fuels: - Natural Gas, LPG, Light Oil, Heavy Oil, OPT1 :
Bituminous Coal, OPT 2 : Anthracite Coal, OPT 3: Coke, OPT 4: Butane, OPT 5:
Wood (Dry), OPT 6: Bagasse by  pressing the Up Arrow Key to select the Fuel
they are working with. When desired fuel is displayed release the key, displayed
fuel is now selected.

After the 30 second countdown the instrument is ready to take Flue, Temperature,
Pressure or Leak Detection readings and Combustion Display 1 will be the first
display.

The 709R will auto power off if no keys have been pressed for 10 minutes and the
CO level is below 15ppm. This feature can be disabled see Appendix F.

The Scroll/Enter Key allows you to scroll through the combustion displays as out-
lined in section 5.3. The Func Key is used to select the 709R function mode and
allows you to set it to be a combustion analyzer, manometer, thermometer, or
combustible leak detector. In addition, the Func Key is used to access the date
and time display. See section 6 for more information on the Func Key.

NOTE:  When selecting oil as fuel be sure to use the optional

oil filter (A773) or readings could become erratic. See
Appendix E for installation instructions. 

Appendix E: ERROR CODES & TROUBLESHOOTING

The 709R analyzer will display certain codes to let you know of a malfunction.

Code

Displayed

Code Definition

Possible Causes

Corrective Action

Flo
Err

Pump not drawing sample at
correct flow rate.

Blockage / kink in flue probe
hose.

Dirty or blocked filter(s).

Worn pump.

Check and rectify.

Replace filter(s).

Return to TPI for service.

InIt

O2

Err

Oxygen sensor failed to ini-
tialize

Flue probe connected to
709R prior to power up.

709R did not purge com-
pletely from last sample.

Worn or defective oxygen
sensor.

Disconnect probe and
restart.

Purge for 20 minutes and
restart.

Return to TPI for sensor
replacement or install new
sensor.*

Try manually initializing the
sensor. See Appendix G.

InIt
CO
Err

Carbon monoxide sensor
failed to initialize.

Flue probe connected to
709R prior to power up.

709R did not purge com-
pletely from last sample.

Worn or defective carbon
monoxide sensor.

Disconnect probe and
restart.

Purge for 20 minutes and
restart.

Return to TPI for sensor
replacement or install new
sensor.*

Try manually initializing the
sensor. See Appendix G.

Lo

bat

Low battery.

Battery needs to be charged. Charge battery. If the battery

won’t hold a charge, replace
the battery.

oFL

Overflow indication. The car-
bon monoxide being mea-
sured is above the instru-
ments capability.

CO being measured is too
high or low.

Remove the CO source and
allow the 709R to purge.

oFL

Overflow indication. The tem-
perature being measured is
outside the maximum mea-
surement capability.

Temperature being measured
is too high or low.

Remove pressure source.

oPEn

Unable to read thermocouple
(temperature).

Temperature probe not con-
nected to input.
Worn temperature sensor.

Connect temperature probe
to analyzer.
Replace temperature probe
or flue probe.

*Sensor replacement requires calibration gas.

Содержание 709R

Страница 1: ... 97005 Tel 503 520 9197 Fax 503 520 1225 www tpi thevalueleader com Test Products International Ltd 342 Bronte Road South Unit 9 Milton Ontario Canada L9T5B7 Tel 905 693 8558 Fax 905 693 0888 709R Combustion Efficiency Analyzer The Value LeaderTM www tpi thevalueleader com ...

Страница 2: ...ocations 5 4 Typical Test Results 6 Function Selection 6 1 Function 1 Thermometer 6 2 Function 2 Manometer 6 3 Function 3 Date Time 6 4 Default Function Combustion Analyzer 7 Saving Data 8 Recalling Data 9 Printing Data Appendix A Specifications Appendix B Calibration Service Appendix C Guarantee Appendix D Installing the Optional A773 Filter Appendix E Error Codes and Trouble Shooting Appendix F ...

Страница 3: ...mical sensors by nature are always active Therefore the time the analyzer is off and not being used must be taken into account when determining sensor life The sensors in your analyzer are war ranted for two years This warranty does not cover sensors damaged through misuse of the analyzer You should keep the batteries of your 709R fresh so power is con stantly being supplied to your sensors The fo...

Страница 4: ...lue Gas Analyzer comes complete with the following stan dard accessories TPI 709R Instrument Rubber Boot A765 Battery Charger A766 1 each Soft Carrying Case A787 Flue Sampling Probe A770 In Line Filter assembly installed on Flue probe A795 Temperature Probe GK11M Pressure Tubing A774 1 each 6 piece Mini Pump Protection Filter Assembly and spare filters A763 Exhaust Spigot removable A764 Instructio...

Страница 5: ...ion chamber also alter the mixture causing decreased efficiency Note 3 The primary way to affect change in combustion efficiency when working with modern gas fired equipment is through gas manifold pressure adjustment Confirm manifold pressures are within maufacturers specifications prior to per forming a combustion test Once manifold pressures are within limits a combus tion test can be performed...

Страница 6: ...WARNING Changing parameters in field calibration mode can adversely affect the operation of your analyzer Follow the instructions carefully 1 Press and hold the Scroll Enter Key down intil FILD CAL is displayed The 709R will cycle through a 30 second countdown then display CAL Air gAS 2 Use the Up Arrow Key to select Air by making it blink Press the Scroll Enter Key to accept the selection 3 709R ...

Страница 7: ...n Press the Scroll Enter Key After the 10 second countdown is complete press the Scroll Enter Key twice 7 709R Display 2000 A r SEt Action Press the Func Key to bypass 8 709R Display Auto oFF En Action Press the Func Key to bypass 9 709R Display CAL StoP go Action Press the Up Arrow Key to select StoP by making it blink and press the Scroll Enter Key to accept the selec tion The 709R will return t...

Страница 8: ... Enter Key to accept the selection 3 709R Display CAL FLo oPEn Action Press the Func Key to bypass 4 709R Display CAL FLo CLos Action Press the Func Key to bypass 5 709R Display 0ppm Co rEF Action Press the Func Key to bypass 6 709R Display 20 9 o2 rEF Action Press the Func Key to bypass 7 709R Display 2000 A r SEt Action If you are not changing the CO alarm point press the Func Key Otherwise pres...

Страница 9: ...t the calculated values of these parameters are out of range Redo combustion test Since these are calculated values the measure values must be within certain levels for these to display If the measured oxygen level is above 19 9 these parameters won t read These parameters might not display or be applicable in some tests Analyzer won t turn off Oxygen and or carbon monoxide levels outside lim its ...

Страница 10: ...on 6 for more information on the Func Key NOTE When selecting oil as fuel be sure to use the optional oil filter A773 or readings could become erratic See Appendix E for installation instructions Appendix E ERROR CODES TROUBLESHOOTING The 709R analyzer will display certain codes to let you know of a malfunction Code Displayed Code Definition Possible Causes Corrective Action Flo Err Pump not drawi...

Страница 11: ...only be switched off if the CO is below 15ppm The instrument has an auto shut off after 10 minutes should no keys have been pressed for this period and as mentioned above that the CO is below 15ppm Should the CO be above 15ppm then the 10 minute auto shut off countdown will not begin till the CO has gone below 15ppm 4 3 Charging The Analyzer Plug the Charger supplied into the charger socket on the...

Страница 12: ...tions for performing tests on commonly encoun tered equipment Remember to consult with the manufacturer of the device under test for specific test information NOTE When selecting oil as fuel be sure to use the optional oil filter A773 or readings could become erratic See Appendix E for installation instructions Appendix B CALIBRATION SERVICE It is recommended that the instrument be calibrated ever...

Страница 13: ...ted the Mini Pump Protection Filter assembly and Temperature Sampling Probe complete with In Line Filter to the Gas Sample Port see 3 2 or 3 4 and the K Type Thermocouple Plug into Thermocouple T1 Socket see 3 4 The GK11M ambient air temperature probe must be connected to the T2 socket refer to figure below for correct hookup WARNING Ensure that the In Line Filter hangs in a vertical position when...

Страница 14: ...percentage Pressing the Up Arrow Key will toggle between Gross Net Efficiency Note Gross Efficiency is used in the USA Combustion Display 3 Displays Carbon Monoxide CO reading in parts per million ppm Displays Oxygen O2 reading in percentage Appendix A SPECIFICATIONS Instrument Operating Temperature Range 14 F to 122 F 10 C to 50 C Battery 1 5V AA size 3 Battery Life 6 Hours Fuels Natural Gas LPG ...

Страница 15: ...y 5 Displays Temperature reading of Channel 1 T1 in degrees Centigrade ºC Displays Temperature reading of Channel 2 T2 in degrees Centigrade ºC Displays the Differential Temperature Diff between ch1 and ch2 in ºC oPEn will be displayed if no K type probe is connected to the thermocouple socket 5 During a combustion test you can save print and send data to a computer See section 7 through 11 Allow ...

Страница 16: ... mode is selected the analyzer will dis play Wait out and bars will change position on the LCD until printing is complete STOR Print Mode If Stor mode is selected Addr will be displayed on the top line along with SA and a location number from 0 to 49 will be flashing on the screen Select the required address location that you wish to print data from by pressing the Up and Down Arrow Keys and press...

Страница 17: ... readings you want to display 4 Press the Scroll Enter key once The 709R will display the date and time screen showing when that reading was stored 5 Use the up and down arrows to scroll through the data that was saved 6 Press the Scroll Enter button and End will be displayed and Yes will flash Press the Scroll Enter button if you are finished recalling data or use the Arrow buttons to select No t...

Страница 18: ...e Up Down Arrow Keys to select Yes and press the Scroll Enter Key 2 Addr and an address location from 0 to 49 will flash on the display Use the Up Down Arrow Keys to select the desired location to store the data and press the Scroll Enter Key To cancel storing information press the Store Key and use the Up Down Arrow Keys to select Yes and press the Scroll Enter Key 3 The data is stored and End wi...

Страница 19: ...n display 1 Combustion Display 1 Default Start Up Display Displays Carbon Monoxide CO reading in parts per mil lion ppm Displays calculated Carbon Dioxide CO2 figure in per centage Displays calculated CO CO2 Ratio figure Function 3 Display Date Time Time Date Month Year COMBUSTION ANALYSIS Continued 5 4 TYPICAL TEST RESULTS Actual test results vary depending on the equip ment under test TPI recomm...

Страница 20: ... temperature tests Function 1 Display Temperature T1 Temperature T2 Temperature Differential Temperature 6 2 Function 2 Manometer Press the Func Key repeatedly to access the thermometer function In this mode the 709R functions like a differential manometer The pump will stop running when in this function 1 Ensure you have Pressure Sampling Tube connected to one or both of the Pressure Ports and th...

Страница 21: ... temperature tests Function 1 Display Temperature T1 Temperature T2 Temperature Differential Temperature 6 2 Function 2 Manometer Press the Func Key repeatedly to access the thermometer function In this mode the 709R functions like a differential manometer The pump will stop running when in this function 1 Ensure you have Pressure Sampling Tube connected to one or both of the Pressure Ports and th...

Страница 22: ...n display 1 Combustion Display 1 Default Start Up Display Displays Carbon Monoxide CO reading in parts per mil lion ppm Displays calculated Carbon Dioxide CO2 figure in per centage Displays calculated CO CO2 Ratio figure Function 3 Display Date Time Time Date Month Year COMBUSTION ANALYSIS Continued 5 4 TYPICAL TEST RESULTS Actual test results vary depending on the equip ment under test TPI recomm...

Страница 23: ...e Up Down Arrow Keys to select Yes and press the Scroll Enter Key 2 Addr and an address location from 0 to 49 will flash on the display Use the Up Down Arrow Keys to select the desired location to store the data and press the Scroll Enter Key To cancel storing information press the Store Key and use the Up Down Arrow Keys to select Yes and press the Scroll Enter Key 3 The data is stored and End wi...

Страница 24: ... readings you want to display 4 Press the Scroll Enter key once The 709R will display the date and time screen showing when that reading was stored 5 Use the up and down arrows to scroll through the data that was saved 6 Press the Scroll Enter button and End will be displayed and Yes will flash Press the Scroll Enter button if you are finished recalling data or use the Arrow buttons to select No t...

Страница 25: ... mode is selected the analyzer will dis play Wait out and bars will change position on the LCD until printing is complete STOR Print Mode If Stor mode is selected Addr will be displayed on the top line along with SA and a location number from 0 to 49 will be flashing on the screen Select the required address location that you wish to print data from by pressing the Up and Down Arrow Keys and press...

Страница 26: ...y 5 Displays Temperature reading of Channel 1 T1 in degrees Centigrade ºC Displays Temperature reading of Channel 2 T2 in degrees Centigrade ºC Displays the Differential Temperature Diff between ch1 and ch2 in ºC oPEn will be displayed if no K type probe is connected to the thermocouple socket 5 During a combustion test you can save print and send data to a computer See section 7 through 11 Allow ...

Страница 27: ...percentage Pressing the Up Arrow Key will toggle between Gross Net Efficiency Note Gross Efficiency is used in the USA Combustion Display 3 Displays Carbon Monoxide CO reading in parts per million ppm Displays Oxygen O2 reading in percentage Appendix A SPECIFICATIONS Instrument Operating Temperature Range 14 F to 122 F 10 C to 50 C Battery 1 5V AA size 3 Battery Life 6 Hours Fuels Natural Gas LPG ...

Страница 28: ...ted the Mini Pump Protection Filter assembly and Temperature Sampling Probe complete with In Line Filter to the Gas Sample Port see 3 2 or 3 4 and the K Type Thermocouple Plug into Thermocouple T1 Socket see 3 4 The GK11M ambient air temperature probe must be connected to the T2 socket refer to figure below for correct hookup WARNING Ensure that the In Line Filter hangs in a vertical position when...

Страница 29: ...tions for performing tests on commonly encoun tered equipment Remember to consult with the manufacturer of the device under test for specific test information NOTE When selecting oil as fuel be sure to use the optional oil filter A773 or readings could become erratic See Appendix E for installation instructions Appendix B CALIBRATION SERVICE It is recommended that the instrument be calibrated ever...

Страница 30: ...only be switched off if the CO is below 15ppm The instrument has an auto shut off after 10 minutes should no keys have been pressed for this period and as mentioned above that the CO is below 15ppm Should the CO be above 15ppm then the 10 minute auto shut off countdown will not begin till the CO has gone below 15ppm 4 3 Charging The Analyzer Plug the Charger supplied into the charger socket on the...

Страница 31: ...on 6 for more information on the Func Key NOTE When selecting oil as fuel be sure to use the optional oil filter A773 or readings could become erratic See Appendix E for installation instructions Appendix E ERROR CODES TROUBLESHOOTING The 709R analyzer will display certain codes to let you know of a malfunction Code Displayed Code Definition Possible Causes Corrective Action Flo Err Pump not drawi...

Страница 32: ...t the calculated values of these parameters are out of range Redo combustion test Since these are calculated values the measure values must be within certain levels for these to display If the measured oxygen level is above 19 9 these parameters won t read These parameters might not display or be applicable in some tests Analyzer won t turn off Oxygen and or carbon monoxide levels outside lim its ...

Страница 33: ... Enter Key to accept the selection 3 709R Display CAL FLo oPEn Action Press the Func Key to bypass 4 709R Display CAL FLo CLos Action Press the Func Key to bypass 5 709R Display 0ppm Co rEF Action Press the Func Key to bypass 6 709R Display 20 9 o2 rEF Action Press the Func Key to bypass 7 709R Display 2000 A r SEt Action If you are not changing the CO alarm point press the Func Key Otherwise pres...

Страница 34: ...n Press the Scroll Enter Key After the 10 second countdown is complete press the Scroll Enter Key twice 7 709R Display 2000 A r SEt Action Press the Func Key to bypass 8 709R Display Auto oFF En Action Press the Func Key to bypass 9 709R Display CAL StoP go Action Press the Up Arrow Key to select StoP by making it blink and press the Scroll Enter Key to accept the selec tion The 709R will return t...

Страница 35: ...WARNING Changing parameters in field calibration mode can adversely affect the operation of your analyzer Follow the instructions carefully 1 Press and hold the Scroll Enter Key down intil FILD CAL is displayed The 709R will cycle through a 30 second countdown then display CAL Air gAS 2 Use the Up Arrow Key to select Air by making it blink Press the Scroll Enter Key to accept the selection 3 709R ...

Страница 36: ...ion chamber also alter the mixture causing decreased efficiency Note 3 The primary way to affect change in combustion efficiency when working with modern gas fired equipment is through gas manifold pressure adjustment Confirm manifold pressures are within maufacturers specifications prior to per forming a combustion test Once manifold pressures are within limits a combus tion test can be performed...

Страница 37: ...mical sensors by nature are always active Therefore the time the analyzer is off and not being used must be taken into account when determining sensor life The sensors in your analyzer are war ranted for two years This warranty does not cover sensors damaged through misuse of the analyzer You should keep the batteries of your 709R fresh so power is con stantly being supplied to your sensors The fo...

Страница 38: ...lue Gas Analyzer comes complete with the following stan dard accessories TPI 709R Instrument Rubber Boot A765 Battery Charger A766 1 each Soft Carrying Case A787 Flue Sampling Probe A770 In Line Filter assembly installed on Flue probe A795 Temperature Probe GK11M Pressure Tubing A774 1 each 6 piece Mini Pump Protection Filter Assembly and spare filters A763 Exhaust Spigot removable A764 Instructio...

Страница 39: ...ocations 5 4 Typical Test Results 6 Function Selection 6 1 Function 1 Thermometer 6 2 Function 2 Manometer 6 3 Function 3 Date Time 6 4 Default Function Combustion Analyzer 7 Saving Data 8 Recalling Data 9 Printing Data Appendix A Specifications Appendix B Calibration Service Appendix C Guarantee Appendix D Installing the Optional A773 Filter Appendix E Error Codes and Trouble Shooting Appendix F ...

Страница 40: ... 97005 Tel 503 520 9197 Fax 503 520 1225 www tpi thevalueleader com Test Products International Ltd 342 Bronte Road South Unit 9 Milton Ontario Canada L9T5B7 Tel 905 693 8558 Fax 905 693 0888 709R Combustion Efficiency Analyzer The Value LeaderTM www tpi thevalueleader com ...

Отзывы: