D. Measuring CO in Flue Gas

1.   Complete steps A and B above.
2. Attach the hose from the flue probe to the Mini Pump 

Protection Filter.

3. Place the shaft of the flue probe into the flue.
4.   706 will measure the CO.  Empty water/air filter as needed.

E. Measuring CO in exhaust

1.   Complete steps A and B above.
2. Attach the hose from the flue probe to the Mini Pump 

Protection Filter. Connect the exhaust probe extension (flexible 
metal tube) to the  end of the flue probe.

3. Place the extension into the exhaust stream.
4.   706 will measure the CO.  Empty water/air filter as needed.

NOTE: The instrument has an auto shut off after 10 minutes should no
keys have been pressed for this period and as mentioned above that
the CO is below 15ppm. Should the CO be above 15ppm then the 10
minute auto shut off countdown will not begin till the CO has gone
below 15ppm.

F. Activating the Backlight

1. Follow the procedure in “A” above.
2. Press and hold down the Func button for 2 seconds.  The back

light will come on and the icon will appear on the LCD.

3. Light will automatically turn off after 20 seconds to preserve 

battery life.

G.  Saving Data

1.   Once you have obtained a reading and want to save it, press 

the Store button once.

2.   The display will show “Stor”. Press the Scroll/Enter button and 

“Addr” and a number from 0 to 49 will flash.

3.   Using the up and down arrows, select the desired location to 

save the readings.

4.  Press the Scroll/Enter key once.  The 706 will store the read

-ings, “End” will be displayed and “Yes” will flash. Press the 
Scroll/Enter button if you are finished printing or use the Arrow
buttons to select No to print again.

Appendix C: ERROR CODES & TROUBLESHOOTING

The 706 analyzer will display certain codes to let you know of a malfunction.

Code

Displayed

Code Definition

Possible Causes

Corrective Action

Flo
Err

Pump not drawing sample at
correct flow rate.

Blockage / kink in flue probe
hose.

Dirty or blocked filter(s).

Worn pump.

Check and rectify.

Replace filter(s).

Return to TPI for service.

InIt
CO
Err

Carbon monoxide sensor
failed to initialize.

Flue probe connected to 706
prior to power up.

706 did not purge completely
from last sample.

Worn or defective carbon
monoxide sensor.

Disconnect probe and
restart.

Purge for 20 minutes and
restart.

Return to TPI for sensor
replacement or install new
sensor.*

Try manually initializing the
sensor. See Appendix G.

Lo

bat

Low battery.

Battery needs to be charged. Charge battery. If the battery

won’t hold a charge, replace
the battery.

oFL

Overflow indication. The car-
bon monoxide being mea-
sured is above the instru-
ments capability.

CO being measured is too
high or low.

Remove the CO source and
allow the 708 to purge.

*Sensor replacement requires calibration gas.

w

w

w

.

G

lo

ba

lT

es

tS

up

pl

y

.c

om

Fi

nd

 Q

ua

lit

Pr

od

uc

ts

 O

nl

in

at

:

sa

le

s@

G

lo

ba

lT

es

tS

up

pl

y.

co

m

Содержание 706

Страница 1: ...4 Beaverton OR 97005 Tel 503 520 9197 Fax 503 520 1225 www tpi thevalueleader com Test Products International Ltd 342 Bronte Road South Unit 9 Milton Ontario Canada L9T5B7 Tel 905 693 8558 Fax 905 693 0888 706 Rev 6 X Analyzers CO Analyzer The Value LeaderTM ...

Страница 2: ...cations Power Supply 4 AA Alkaline Batteries Sensor Long Life Electrochemical Range O to 9 999 ppm Resolution 1ppm 0 to 9999ppm 0 001 9999ppm Accuracy 5 ppm or 5 of reading whichever is greater Warm up Approx 30 seconds Response time 90 of reading 120 seconds Duty Cycle 30 minutes Battery Life Approx 30 hrs Size 3 75 x 8 0 x 3 0 95 x 203 x 76 mm Weight 1 1 lbs 500g Probe Length 9 230mm Tube Length...

Страница 3: ...able General Description The 706 is an advanced state of the art carbon monoxide CO gas analyzer designed to provide part per million readings from 0 10 000 ppm The 706 is capable of performing ambient as well as flue gas checks with its hot air probe and internal pump The 706 is utilizes an advanced long life electrochemical sensor that can be field calibrated and has a low replacement cost Featu...

Страница 4: ...on of your analyzer Follow the instructions carefully 1 Press and hold the FUNC and Down Arrow keys for approximately 5 seconds until a beep is heard Press and hold the Scroll Enter Key down until FILD CAL is displayed The 706 will cycle through a 30 second countdown then display CAL Air gAS 2 Use the Up Arrow Key to select Air by making it blink Press the Scroll Enter Key to accept the selection ...

Страница 5: ... changing the auto power off mode press the Func Key Otherwise press the Up and Down Arrow Keys to enable En or disable dis the auto off function Factory default is enabled En Press the Scroll Enter Key to accept the selection 8 706 Display CAL StoP go Action Press the Up Arrow Key to select StoP by making it blink and press the Scroll Enter Key to accept the selec tion The 708 will return to norm...

Страница 6: ...e Func Key Otherwise press the Up and Down Arrow Keys to enable En or disable dis the auto off function Factory default is enabled En Press the Scroll Enter Key to accept the selection 8 706 Display CAL StoP go Action Press the Up Arrow Key to select StoP by making it blink and press the Scroll Enter Key to accept the selec tion The 708 will return to normal operation Operating Instructions A Powe...

Страница 7: ...ecall button once 2 The display will show Stor and Cal Use the Arrow buttons to select the desired function Stor Selecting Stor allows you to display the information in a memory location 1 Press the Recall button once The display will show Stor and Cal Using the Arrow buttons select Stor by making it blink Press the Scroll Enter button 2 Addr will be displayed and a number from 0 to 49 will flash ...

Страница 8: ...ode If Cal print mode is selected the analyzer will display Wait out and bars will change position on the LCD until printing is com plete After printing is complete End will be displayed and YES or no will blink Selecting YES will return the 706 to normal operation selecting no will return the 706 to the print selection screen so you can print again Use the Up and Down Arrow Keys to make the selec...

Страница 9: ...LCD until printing is complete CAL Print Mode If Cal print mode is selected the analyzer will display Wait out and bars will change position on the LCD until printing is com plete After printing is complete End will be displayed and YES or no will blink Selecting YES will return the 706 to normal operation selecting no will return the 706 to the print selection screen so you can print again Use th...

Страница 10: ...Stor and Cal Use the Arrow buttons to select the desired function Stor Selecting Stor allows you to display the information in a memory location 1 Press the Recall button once The display will show Stor and Cal Using the Arrow buttons select Stor by making it blink Press the Scroll Enter button 2 Addr will be displayed and a number from 0 to 49 will flash 3 Using the up and down arrows select the ...

Страница 11: ...oll Enter button and Addr and a number from 0 to 49 will flash 3 Using the up and down arrows select the desired location to save the readings 4 Press the Scroll Enter key once The 706 will store the read ings End will be displayed and Yes will flash Press the Scroll Enter button if you are finished printing or use the Arrow buttons to select No to print again Appendix C ERROR CODES TROUBLESHOOTIN...

Страница 12: ...e Func Key Otherwise press the Up and Down Arrow Keys to enable En or disable dis the auto off function Factory default is enabled En Press the Scroll Enter Key to accept the selection 8 706 Display CAL StoP go Action Press the Up Arrow Key to select StoP by making it blink and press the Scroll Enter Key to accept the selec tion The 708 will return to normal operation Operating Instructions A Powe...

Страница 13: ...tion mode can adversely affect the operation of your analyzer Follow the instructions carefully 1 Press and hold the FUNC and Down Arrow keys for approximately 5 seconds until a beep is heard Press and hold the Scroll Enter Key down until FILD CAL is displayed The 706 will cycle through a 30 second countdown then display CAL Air gAS 2 Use the Up Arrow Key to select Air by making it blink Press the...

Страница 14: ...anced state of the art carbon monoxide CO gas analyzer designed to provide part per million readings from 0 10 000 ppm The 706 is capable of performing ambient as well as flue gas checks with its hot air probe and internal pump The 706 is utilizes an advanced long life electrochemical sensor that can be field calibrated and has a low replacement cost Features include IR printer port to print hard ...

Страница 15: ... the CO Sensor Appendix F CO Chart Specifications Power Supply 4 AA Alkaline Batteries Sensor Long Life Electrochemical Range O to 9 999 ppm Resolution 1ppm 0 to 9999ppm 0 001 9999ppm Accuracy 5 ppm or 5 of reading whichever is greater Warm up Approx 30 seconds Response time 90 of reading 120 seconds Duty Cycle 30 minutes Battery Life Approx 30 hrs Size 3 75 x 8 0 x 3 0 95 x 203 x 76 mm Weight 1 1...

Страница 16: ...706 Rev 6 X Analyzers CO Analyzer The Value LeaderTM www tpi thevalueleader com www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Отзывы: