Press and hold down the 

HOLD

 button for approximately 8 seconds until “CAL” is displayed. “CAL” will 

display for approximately 2 seconds and then the 310C will return to normal operation. Calibration is complete.  
 
Note: If the temperature reading was not within 23°F to 41°F (-5°C to 5°C) when step 3 was performed no 
change in calibration occurred. The 310C was designed with this feature to prevent improper calibration. 
 