Motion Limit Controls

 

 

Function 

The  Mirage  Hydraulic  Patient  Chair  does  not  use  conventional  mechanical  or  mercury  limit 
switches.    Motion  of  the  chair  base  and  backrest  move  variable  resistances  (potentiometers) 
making it possible for the control circuitry to continuously know the exact position of the base and 
backrest, not only when they hit the end of their travel.  Adjustments for all four actions (seat raise/ 
lower, backrest raise/recline) are located on the control PCB. 
 

Adjustment 

To set a specific limit for the mirage operatory chair, perform the following instructions. Remove the 
pump cover from the chair. Locate the slider switch labeled S1. The switch is located on the lower left 
corner of the main PCB board. This will put the chair into service mode. Use the touch pad or foot control 
to move the chair to the desired position. See below for detailed instructions on setting each limit. A 
diagram of the PCB board is on the following page.

 

Seat Lower Limit 

Using the manual controls, set seat height to the position where the lower limit is set.  Press Lift DO on the 
main PCB board to program. Reset the Slide Switch on PCB to the “off” position. Use the touch pad or foot 
control to move the chair up, and then use the lift down on the touch pad or foot control to verify that the 
new limit is set. 

Seat Height Limit 

Using the manual controls raise the seat height fully. At the physical top of travel, the motor will 
continue to run, but upward motion will stop (oil bypass channel has opened in cylinder).  Lower 
the seat height slightly from this point and press LIFT UP button on Control PCB. Reset the Slide 
Switch on PCB board to the “off” position. Then use the touch pad or foot control to lower the 
chair, raise the chair and verify the desired height limit is set.  

 

Backrest Recline Limit 

Using the manual controls, Recline the backrest fully.  Raise the backrest slightly from this point 
and press the BACK DN button on the main PCB board. Reset the Slide Switch on PCB board to 
the “off” position. Then raise the back rest and then lower it to verify the desired setting is set.  

 

Backrest Up Limit 

Using the manual controls, raise the backrest fully. Then lower the backrest slightly from this point. Press 
the backup button on the main PCB board. Reset the Slide Switch on PCB board to the “off” position. 
Then use the foot control disc to lower the backrest. Raise the backrest to verify the desired height limit is 
set. 

Содержание Mirage 2.0 Hydraulic

Страница 1: ...1 www tpcdental com Mirage 2 0 Hydraulic Patient Chair Installation Operation User Manual TPC 851 S Lawson St City Of Industry CA 91748 P626 810 4337 Fax 626 810 4245...

Страница 2: ...n Contents Page GENERAL INFORMATION 3 Installation and coordination 5 Installing dental chair upholstery 7 Overview of controls 11 Calibrating Chair Motors 13 CLEANING DISINFECTING GUIDE 14 Warranty 1...

Страница 3: ...number and date of manufacture This label is located on the left side cover of the dental chair The dental chair is marked with a Risk Class symbol This label is located on the left side cover of the...

Страница 4: ...et jack or forklift to move the dental chair Do not attempt to move stacked dental chairs Only store the dental chair in dry cool place The dental chair is marked with a main electrical grounding poin...

Страница 5: ...s 1 Remove the plastic covers from the base of the dental chair Slide the back cover straight back to remove it from the chair base 2 Remove the 4 shipping bolts that secure the chair to the pallet Tw...

Страница 6: ...ir These are going to be used as a lift point Do not use the chair armrest as lift points 4 After the chair is placed in the installation location Install the leveling lugs in the holes that were used...

Страница 7: ...s from the backrest mount by removing the two 14mm bolts 6 Secure the backrest support to mounting hub using the 4 6mm Allen bolts You don t need to remove the plastic cover Gently slide the backrest...

Страница 8: ...f the nut can slide up and down with little resistance in the back rest frame 4 Once the space is adjusted remove the bolts with the nuts still attached and try to keep the spacing the same 5 Place th...

Страница 9: ...s bracket shown in figure 2 to the upholstery 10 Mount the s bracket as shown in figure 3 11 Attach the two supplied Allen screws to the chair seat cushion frame shown in figure 3 Incorrect Mounting t...

Страница 10: ...t to right 13 locate the Allen screws lock washer and washers shown in figure 2 14 Once the bracket is securely in place you may fasten the toe of the upholstery to the frame as shown in figure 3 15 A...

Страница 11: ...11 www tpcdental com 1 Place the plastic cover over the headrest cushion 2 Attach the headrest cushion to the headrest stem using 3 2 5MM Allen screws Bag E...

Страница 12: ...rol 3 Preset Operation Manually move the chair to the desired position Press and hold the pre set switch 1 or 2 on the foot control to program After 5 seconds you will hear a confirmation beep 4 Autom...

Страница 13: ...tal com Cuspidor Cup Fill Cuspidor Bowl Rinse Overview of optional touchpad controls Chair Base down Chair Base up Chair back down Chair Back UP Preset 1 2 and 3 Last Position Auto Return Light On Off...

Страница 14: ...eath Armrest Rotation Either armrest can be rotated outward by grasping the end portion of armrest pulling upward and turning Headrest Adjustments Height of headrest can be adjusted simply by pulling...

Страница 15: ...ram Reset the Slide Switch on PCB to the off position Use the touch pad or foot control to move the chair up and then use the lift down on the touch pad or foot control to verify that the new limit is...

Страница 16: ...ugh two complete cycles of movement The first cycle is the maximum travel of the chair The second cycle will deduct about 10 of the first measured movement from cycle 1 Once this procedure is complete...

Страница 17: ...17 www tpcdental com Wiring schematic...

Страница 18: ...TV2 Base down Flow Control TV3 Back up flow control TV4 Back down flow control When adjusting the flow control valves counterclockwise turns will open increase flow Turning the valve clockwise will c...

Страница 19: ...ating and is in the UP position On Check the two 10 Amp fuses to the right of the switch Disconnect the main power cord when checking the fuses These fuses are 10Amp 250v 5X20mm glass fuses Replace if...

Страница 20: ...disinfectant to soak in or dry on the surface wipe off with soap and water PLEASE KEEP LIQUIDS AWAY FROM ELECTRICAL CONNECTIONS Most electrical control switches ARE NOT hermetically sealed When a swit...

Страница 21: ...ervoir Dental Chair does not go low or high enough Program limits Re program limit switch settings Unable to program preset position 1 2 and 3 Chair is in program mode and not operational mode Put cha...

Страница 22: ...er normal wear or stains on surface finish The guarantee does not cover damage resulting from improper installation misuse or accidents incurred in shipping and handling All claims against the freight...

Отзывы: