Enter a test label.
Use the [
▲
] and [
▼
] keys to
scroll and press [Next]
to select.
Select empty space to store
label and go to the next screen.
6
Place the oil sample contained in a
Diagomar Plus bottle on the magnetic
rack.
Beware: in order to separate
iron from particulate iron and ensure
test accuracy, the bottle must remain
on the rack for 5 minutes prior to
testing.
Press the [Iron] key.
Up to 7 vials containing
reagents can be prepared in
advance as soon as they are
used within 2 hours.
Press any of the 4 front panel
keys to turn the Rapid Optical
Oil Tester on.
Then press the [Test] key.
Follow the test instructions
displayed by the instrument.
Press the [
▼
] key for more
instructions.
Close the vial with a new cap
and flip it over once
smoothly
to mix the contents.
After mixing, it is normal
that the reagents remain in
two phases.
When the instrument memory
is full, a “History Full” message
shows. Press [Go] to confirm test
(result will not be stored) or
[Abort] to cancel.
Using 1 clean syringe per
reagent, inject into the vial:
5 ml of Solution IT1
5 ml of Solvent IT2
5 ml of Reagent IT3
2
5
9
3
7
10
4
8
14
DO NOT SHAKE
Iron
RUN TEST
LUBMARINE
Zero
5° Homogenize mix gently
6° Insert vial into Tester
7° Press ZERO
Run Iron Test
Next
Enter Test id
……
Run Iron Test
LUBMARINE
Test
Conf
Hist
RAPID OPTICAL
OIL TESTER
Abort
Go
History Full
Iron Test
With Rapid Optical Oil Tester