background image

 TIC-LF470B

 

 

2

 

Model LF470 Detectors 

Fluid pressure: 

-15 to 150 psi, or -1.0 to 10 bar   

(-0.1 to 1 MPa)

 

Principal materials:

 

Measurement tube – 

Alumina ceramics 

Electrodes

 

– 

Platinum 

Pipe connection port: 

Standard

 – 

316 stainless steel 

Option  

 Ti (titanium), 
Polyvinyl chloride (shock-res 

Ta (Tantalum) for grounding plate, 
Polyvinyl chloride (shock-res 
Pt-Ir grounding plate, 
Hastelloy C (Equivalent) 

Dedicated preformed cables:

 

Signal cable

 

– 2-wire shielded chloroprene 

sheathed cable

 

                      Overall length: 0.28 inch (7mm) 
           Length: 0.20 inch (5m) 

Excitation cable 

 3-wiren chloroprene sheathed 

cable 

Overall length: 0.28 inch (7mm) 

  Length: 

0.20 inch (5m)

 

Coating: 

phthalic acid resin coating pearl-gray 

colored 

Weight:

 approximately 2.2 lb (1.0 kg) 

 

               

(for each size excluding cables) 

 

 

Model LF612 converters 

Input signals

 

Analog signal 

 

the voltage signal from detector, 

proportional to process flow rate (for LF612 

separate type converter). 

Digital input DI (opt.) 

Signal type:    20 to 30Vdc voltage signal 

Input resistance: 2.7k

 

Number of inputs: one point 

DI function 

— One of the following functions can be 

assigned to the optional DI signal. 

Range switching 

— Selects either the higher or lower 

range in the unidirectional or bidirectional 2-range 

setting.

 

Totalizer control 

— Starts and stops the built-in 

totalizer. 

Fixed-value outputs 

—Outputs fixed-values for 

current and pulse outputs. 

Zero adjustment 

— Executes zero adjustment 

(on-stream at zero flow rate). 

Output signals 

Current output: 

4–20mAdc (load resistance 0 to 750

Note: 

The current output cannot be used with the 

PROFIBUS-PA communication. 

Digital outputs 

— One point (std.) and one more 

point is optionally available as follows. 

Digital output DO1 (std.): 

Output type: Transistor open collector 
Number of outputs: One point 
Output capacity: 30Vdc, 200mA maximum 

Digital output DO2 (opt.): 

Output type:   

Solid-state relay output (non polarity) 

Number of outputs: One point 
Output capacity: 150Vdc, 150mA maximum   

or 150Vac (peak to peak), 100mA maximum 

DO1 and DO2 functions 

— One of the following 

functions can be assigned to DO1 (std.) and/or 
DO2 (opt.) 

• 

Pulse output (available only for DO1, DO2) 

Pulse rate: 3.6 to 36,000,000 pulses/hr (DO1) 

3.6 to 360,000 pulses/hr (DO2) 

(Over 3,600,000 pulses/hr, auto-setting) 

Pulse width: 0.5 to 500ms (but less than half of 
the period for 100% flow rate) 

Note:

 The same and simultaneous

 

 pulse is not 

available between DO1 and DO2.) 

• 

Multi-range selection outputs (Note 1) 

• 

High, High high, Low, and/or Low low alarm 

outputs (Note 2) 

• 

Empty pipe alarm output 

• Digital Output Active Status (DO1 and DO2) 

(Note 2) 

• Preset count output 
• Converter failure alarm output 

Note 1:

 Two outputs (DO1 and DO2) are needed for 

4-range switching and forward/reverse 2-range 
switching. 

Note 2:

Normal Open (default set) or Normal Close is 

selected for alarm outputs when programming. 

When power failure occurs, unit will be fault to 
Normal Open.

 

 

Содержание Electromagnetic Flowmeter LF470/LF612

Страница 1: ...icator can be used to communicate with the flowmeter from a remote place PROFIBUS PA 3 interface is available as an option 1 Please refer to the document TIC LF232 2 HARTprotocol Highway Addressable Remote Transducer is a communication protocol for industrial sensors recommended by the HCF HARTCommunication Foundation 3 PROFIBUS is the communication protocol for factory and process automation that...

Страница 2: ...for current and pulse outputs Zero adjustment Executes zero adjustment on stream at zero flow rate Output signals Current output 4 20mAdc load resistance 0 to 750Ω Note The current output cannot be used with the PROFIBUS PA communication Digital outputs One point std and one more point is optionally available as follows Digital output DO1 std Output type Transistor open collector Number of outputs...

Страница 3: ...nal Corp Conditions when power fails Parameter setting values are stored in non volatile memory and the values will be restored when the power returns to normal condition The outputs and display will remain as follows when power fails Current output 0mAdc Digital output OFF LCD display No display PROFIBUS No communication Power supply One of the following can be selected 100 to 240Vac 50 60Hz std ...

Страница 4: ...39 86 2 φ0 24 φ6 hole 4 φ0 20 φ5 hole Detector mounting plate 4mm thick Mountingscrews 1 18 30 1 57 40 Figure4 LF470 Mounting Board 矢視A 信号ケーブル グランド 電源ケーブルグランド 封止栓 励磁ケーブルグランド 電流出力ケーブルグランド 接地端子 M4 取付金具 4 φ 11 取付金具 82 36 156 13 151 178 200 151 74 0 3 25 LCD 表示器 赤外線スイッチ Excitation cable ground 5 94 151 5 94 151 2 91 74 4 φ0 43 φ11 Attachment LCD display 7 01 178 7 87 200 IR Switch Cable ground 6 14 15...

Страница 5: ...l input 20 30Vdc Power supply Instrument panel ordered separately Figure 6 Separate type LF470 LF612 flowmeters wiring Diagram Table 1 LF612 Converters Signal Table Symbol Description Cable L1 L2 Power supply Power cable CVV GND Ground for arrester FG Frame ground DI Digital Input 20 30Vdc DO1 Digital Output 1 DO2 Digital Output 2 COM Signal Common for DI DO1 DO2 Current Output 4 20mAdc or PROFIBU...

Страница 6: ...60 4 482 Table 3 Meter size and Flow Velocity US unit Unit L min Meter size mm 0 3m s 1m s 3m s 10m s 2 5 0 08835 0 2945 0 884 2 945 4 0 2262 0 7540 2 262 7 540 6 0 5088 1 6967 5 090 16 967 About establishment environment Do not store or install the flowmeter Where there is direct sunlight Where excessive vibration or mechanical shock occurs Where high temperature or high humidity conditions exist...

Страница 7: ...l chloride shock resistance Pt Ir for grounding plate H Hastelloy C Equivalent Packing used between main body and pipe connection adapter C Acid resistant ceramic D Alkali resistant ceramic G Ammonia resistance Mounting A Not provided B Provided 304 stainless steel Flow and calibration velocity range A 1 0 to 32 8 ft s standard range calibration B 1 0 to 32 8 ft s specified range calibration Cable...

Страница 8: ...ting BNP material SUS304 Accessory for pipe installation BNP material SUS304 1 2 Digital input output Digital output points 1 DO1 Digital output points 2 DO1 DO2 Digital input point 1 DI 1 2 Current output and Communication function Current output HART communication PROFIBUS communication Current output is not usable 1 2 3 Power supply 100Vac 240Vac 50 60Hz 24Vdc 110Vdc E Instruction manual Englis...

Отзывы: