Torrey Pines Scientific, Inc.
33
send: le<CR>
returned: ok <CR><LF>
Utility Commands
Command:
#m
Function:
Return all calibration values
Description: When the command #m<CR> is received by the SC20 unit, the unit will return 4 values
separated by commas and terminated by <CRLF>.
The values and order are:
Calibration_Temp_Lo, Measured_Temp_Lo, Calibration_Temp_Hi, Measured_Temp_Hi<CRLF>
Calibration_Temp_Lo: Set Point temperature when calibration at low temperature was done.
Measured_Temp_Lo: Measured temperature entered when calibration at low temperature was done.
Calibration_Temp_Hi: Set Point temperature when calibration at high temperature was done.
Measured_Temp_Hi: Measured temperature entered when calibration at low temperature was done.
Example – return the stored calibration values
send: #m<CR>
returned: 37, 36.2, 70, 71.1<CR><LF>
Command:
#F
Function:
Reset all Calibration Values to Factory Set Values
Description: When the command #F<CR> is received by the SC20 unit, the unit will reset and store all
calibration values to those measured and set in the factory prior to shipment.
Example – reset cal vals back to factory
send: #F<CR>
returned: ok<CR><LF>
Command:
#Z
Function:
Reset Unit to Default Configuration
Description: When the command #Z<CR> is received by the SC20 unit, the unit will reset all non-volatile
memory to default factory values.
Example – reset Unit
send: #Z<CR>
returned: Unit Reset<CR><LF>