![TORREY PINES SCIENTIFIC EchoTherm SC20 Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/torrey-pines-scientific/echotherm-sc20/echotherm-sc20_operating-manual_1134656014.webp)
Torrey Pines Scientific, Inc.
14
stopped, the last timer value is “frozen” on the screen until either the Up or Down buttons
are pressed. This enables the user to capture the last value.
The Menu also includes Timer Options to control the features of the timer. The audible
alarm may be enabled or disabled when the timer reaches zero. If the “Auto-Off” feature is
disabled (“Auto-Off: no”), the unit will continue to control the plate to the set point
temperature when the timer reaches zero. If the “Auto-Off” feature is enabled (“Auto-
Off:yes”), the unit will go into “Idle Mode” when the timer reaches zero. When the unit is in
Idle Mode, no power will be directed to the plate—the plate is essentially “turned off” and
will naturally return to room temperature. See the Idle Mode section of this manual for
more details.
Figure 2:
Menu Map to Set and Start the Timer