Clearing Calibration 
 
Before setting calibration, always clear the existing calibration.  To clear calibration: 
 

1.  Hold the TEMP key for 5 seconds until the display says CALIBRATE? on the top 

line and ENTER/CANCEL on the bottom line of the display. 

2. Depress ENTER 
3.  Then depress ERASE.  The display will now say ERASE CAL? On the top line 

and ENTER/CANCEL on the bottom line of the display. 

4. Depress ENTER 

 
The existing calibration is now erased. 
 
Single Point Calibration 
1- 

Set the unit to the target temperature to be calibrated.  Measure the chamber 

 

temperature using your laboratory standard.  Be sure to allow the chamber and 

 

your thermometer time to equilibrate at the target temperature you desire. 

 
2- 

Push and hold the TEMP key for five seconds until the display says 

 

CALIBRATE? on the top line of the display and ENTER/CANCEL on the bottom 

 

line.   

 
3- Depress 

ENTER. 

  
4-  

Enter the value onto the display as read from your temperature standard and 
depress ENTER.     

 

 

Single point calibration is now complete and will be held in memory until changed. 
 
Two Point Calibration 
 
For two point calibration, one calibration temperature must be above 37.0 ºC and the 
other must be below 37.0º C.  For best results, use temperatures around 10.0 ºC and 
60.0º C.  Follow the calibration procedure as for a single point.  Then repeat the 
procedure at the second point.  When you calibrate the second point, the calibration of 
the first point is not affected.  However, all other temperatures will adjust to a “straight 
line” through the two calibration points.   

 

NOTE:

 

Measuring the chamber temperature is easy to do, but not necessarily 

accurately.  Generally speaking, glass thermometers found around a lab are not 
accurate enough or readable enough to do a good job.  Good, certified glass 
thermometers are available from lab supply dealers as are fairly accurate digital 
thermometers

.  

 
 
12 

Содержание EchoTherm IN35-12VDC

Страница 1: ...TOR MODELS IN35 AND IN45 12VDC DOCUMENT NUMBER IN35 03 Revised 15 February 2012 TORREY PINES SCIENTIFIC INC 2713 Loker Ave West Carlsbad CA 92010 Telephone 760 930 9400 Fax 760 930 9480 E Mail info to...

Страница 2: ...F ITEMS 3 IV CAUTIONS 3 4 V SET UP 4 5 VI GENERAL DESCRIPTION 5 OPERATION VII FRONT AND REAR PANEL 6 7 VIII SIMPLE OPERATION 8 IX SIMPLE TEMP TIMER RAMP AND AUTO OFF SETTINGS 8 9 10 X PROGRAMMABLE OPE...

Страница 3: ...eturned For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss Torrey Pines Scientific will not be liable for damage resulting fro...

Страница 4: ...n toward the front of the chamber where it can easily be cleared GENERAL The IN35 and IN45 are designed for laboratory use and for use only as instructed in this manual If the instrument is used in a...

Страница 5: ...hold temperatures accurately and reproduce temperatures exactly The chamber is molded ABS plastic and the racks are electro polished steel Since the units are Peltier driven there is a fan in the cham...

Страница 6: ...ture in the center of the top line of the display and the timer with all zeros across the bottom line The display is a two line 16 digit alphanumeric back lighted LCD When a target temperature has bee...

Страница 7: ...RS232 and a reorder label The I O port will be discussed later in the chapter on the RS232 The label tells the model number serial number operating voltage of the unit power consumption and the addres...

Страница 8: ...ove the settable digit to the left Now use the UP ARROW or DOWN ARROW to scroll to the temperature setting wanted for that digit Use the RIGHT ARROW to move to the next digit to the right Use the UP A...

Страница 9: ...or RIGHT ARROW 3 When the value is set press ENTER The ramp rate is now set SETTING TIMER The timer is a count down timer that can be set to a maximum of 99 days 59 hours 59 minutes and 59 seconds It...

Страница 10: ...R will erase the programs and the display will show PLEASE WAIT The programs will then be erased This erases ALL programs in memory Hitting CANCEL will cancel the instruction To change or erase one pr...

Страница 11: ...m hit the UP ARROW 7 After the last program step data has been entered press ENTER The display will now show CYCLE 01 This gives the user the choice of having the program just entered cycle one or mor...

Страница 12: ...NCEL on the bottom line 3 Depress ENTER 4 Enter the value onto the display as read from your temperature standard and depress ENTER Single point calibration is now complete and will be held in memory...

Страница 13: ...the RS232 interface capable of doing this The instructions for connection and use follow RS232 INTERFACE SPECIFICATION Parameters 2400 baud 8 data bits 1 stop bit no parity No handshake hardware or s...

Страница 14: ......

Отзывы: