NOTE: The timer will count down to zero. At zero, the audible alarm will sound
once a second for one minute. Also, as the timer starts to sound, the display for
the timer will start to count up letting the user know how much time has passed
since the timed even ended. The audible alarm can be turned off any time during
the first minute by pressing CANCEL.
X. PROGRAMMABLE OPERATION
The IN35 and IN45 have the ability to store three programs in memory for instant recall
and use at any time. Each program has ten steps maximum where one step is a
temperature, time and ramp function (if wanted). At the end of each program there is a
cycle step that can be selected which will repeat the program up to 99 times. These
programs are stored in CMOS and cannot be lost by a power failure or by turning off the
unit. However, if power fails during a program run, when it returns, the unit will have
lost its instructions and will be sitting with the same program step as when power failed.
Some rules for programming:
To erase all programs in memory, depress and hold the EDIT key for about 5
seconds. The display will show ERASE PROGRAMS, ENTER/CANCEL. Hitting
ENTER will erase the programs and the display will show PLEASE WAIT... The
programs will then be erased. This erases ALL programs in memory. Hitting
CANCEL will cancel the instruction. To change or erase one program the user
must go into the program and remove each instruction.
The timer can be used in two ways in a program. The timer can be used with a
ramp function or without a ramp function. When the timer is used with a ramp
function, the timer does not start to count down until the target temperature is
reached. When the timer is used without a ramp function, the timer starts to
count down as soon as the program starts. If the user wants a timed event to
start after the target temperature is reached, but does not want to wait for a
ramped approach to the target temperature to take place, set the ramp at 99.9. In
almost all cases, this ramp rate is faster than the unit can approach the target
temperature without a RAMP, and it will appear as though no ramp rate is being
used. The timer will start to count down after the target temperature is reached.
RUNNING A PROGRAM STORED IN MEMORY
1
Press RUN. The display will show PROGRAM 1.
2
If program 1 is to be run press ENTER. If program 1 is not the choice, use the
UP ARROW to scroll to program 2 or 3. When the program is chosen, press
ENTER and the program stored in that site will start to run.
10