background image

 

 

 

 
 

OPERATING MANUAL 

 

EchoTherm™ 

PROGRAMMABLE DIGITAL CHILLING INCUBATOR 

MODELS IN35 AND IN45-12VDC 

 
 
 
 
 
 
 
 

DOCUMENT NUMBER IN35-03 

Revised 15 February 2012 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

TORREY PINES SCIENTIFIC, INC. 

2713 Loker Ave. West 

Carlsbad, CA 92010 

 

Telephone: (760)-930-9400 

Fax: (760)-930-9480 

E-Mail: [email protected] 

Web Site: www.torreypinesscientific.com 

Содержание EchoTherm IN35-12VDC

Страница 1: ...TOR MODELS IN35 AND IN45 12VDC DOCUMENT NUMBER IN35 03 Revised 15 February 2012 TORREY PINES SCIENTIFIC INC 2713 Loker Ave West Carlsbad CA 92010 Telephone 760 930 9400 Fax 760 930 9480 E Mail info to...

Страница 2: ...F ITEMS 3 IV CAUTIONS 3 4 V SET UP 4 5 VI GENERAL DESCRIPTION 5 OPERATION VII FRONT AND REAR PANEL 6 7 VIII SIMPLE OPERATION 8 IX SIMPLE TEMP TIMER RAMP AND AUTO OFF SETTINGS 8 9 10 X PROGRAMMABLE OPE...

Страница 3: ...eturned For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss Torrey Pines Scientific will not be liable for damage resulting fro...

Страница 4: ...n toward the front of the chamber where it can easily be cleared GENERAL The IN35 and IN45 are designed for laboratory use and for use only as instructed in this manual If the instrument is used in a...

Страница 5: ...hold temperatures accurately and reproduce temperatures exactly The chamber is molded ABS plastic and the racks are electro polished steel Since the units are Peltier driven there is a fan in the cham...

Страница 6: ...ture in the center of the top line of the display and the timer with all zeros across the bottom line The display is a two line 16 digit alphanumeric back lighted LCD When a target temperature has bee...

Страница 7: ...RS232 and a reorder label The I O port will be discussed later in the chapter on the RS232 The label tells the model number serial number operating voltage of the unit power consumption and the addres...

Страница 8: ...ove the settable digit to the left Now use the UP ARROW or DOWN ARROW to scroll to the temperature setting wanted for that digit Use the RIGHT ARROW to move to the next digit to the right Use the UP A...

Страница 9: ...or RIGHT ARROW 3 When the value is set press ENTER The ramp rate is now set SETTING TIMER The timer is a count down timer that can be set to a maximum of 99 days 59 hours 59 minutes and 59 seconds It...

Страница 10: ...R will erase the programs and the display will show PLEASE WAIT The programs will then be erased This erases ALL programs in memory Hitting CANCEL will cancel the instruction To change or erase one pr...

Страница 11: ...m hit the UP ARROW 7 After the last program step data has been entered press ENTER The display will now show CYCLE 01 This gives the user the choice of having the program just entered cycle one or mor...

Страница 12: ...NCEL on the bottom line 3 Depress ENTER 4 Enter the value onto the display as read from your temperature standard and depress ENTER Single point calibration is now complete and will be held in memory...

Страница 13: ...the RS232 interface capable of doing this The instructions for connection and use follow RS232 INTERFACE SPECIFICATION Parameters 2400 baud 8 data bits 1 stop bit no parity No handshake hardware or s...

Страница 14: ......

Отзывы: