Transducer Control 5 User Manual (RWT3746IM)
Revision 5, December 2016 - Page 25
Error
Error flags indicate that a system fault has been detected. System faults cannot be
corrected by the user. If a fault condition does occur, try power cycling the
transducer. If the fault does not clear the transducer will have to be returned for
service.
Electronics Fault (Firmware Version 5)
An electronics fault has been detected.
Head Disconnect (Firmware Version 3/4)
Please refer to the previous section for a description of this error. In firmware
version 5 this error was reclassified as a warning as it is a condition that is
correctable by the user.
SAW Error
A SAW error indicates that the shaft attached SAW devices are not operating
within normal parameters.
Failed Temperature Sensor
A failed temperature sensor error indicates that the transducers internal shaft
temperature sensor has malfunctioned. When the temperature sensor fails the
transducer will no longer compensate the torque readings for changes in
temperature, which will cause a loss in accuracy.
Irregular Internal Voltage
An irregular internal voltage error indicates that main processor has detected that
one of the internal voltage rails is operating outside of normal parameters.
Data Configuration Invalid
A data configuration invalid error indicates that main build configuration has been
corrupted.
Calibration Invalid
A calibration invalid error indicates that the calibration data has become
corrupted.
SAW Parameters Invalid (RWT Transducers)
A SAW parameters invalid error indicates that an internal configuration structure
has become corrupted.
ORT Parameters Invalid (ORT Transducers)
An ORT parameters invalid error indicates that an internal configuration structure
has become corrupted.
Содержание ORT230-C
Страница 30: ......
Страница 31: ...Transducer Control 5 User Manual Revision 5 December 2016...