35
The LED of DB-67C will indicate charging status:
Red ON
: Charging.
Green ON
: Charging completed.
Green flashing : Ni-MH BT-67Q battery pack is not installed correctly.
Red flashing
: Ni-MH BT-67Q battery pack protection feature is working automatically.
RL-100 2S can be used in this state.
The instrument has a protection feature which works when nickel hydride batteries are
overcharged or when the batteries are under a high or low temperature (+70°C or higher, or 0°C or
lower) state. In such a case, charging will stop automatically to protect nickel hydride batteries.
Recharging should be performed in a room temperature ranging from +10°C to +40°C.
Always use the AC/DC converter provided with the product.
1) The Ni-MH BT-67Q rechargeable battery can be charged while using the laser.
2) The Ni-MH BT-67Q rechargeable battery can be charged when the battery holder is
removed from the instrument. This allows the option of alternately using two battery
packs to always maintain a fully charged pack.
3) The Ni-MH BT-67Q rechargeable battery can be removed from the DB-67C battery
holder and four “D” cell alkaline batteries can be installed.
4)The DB-67 dry cell battery holder cannot be used to charge the BT-67Q Ni-MH battery
pack. Use the DB-67C charging battery holder instead.