If  the  chair  has  a  1

st

  generation  circuit  board,  remove  the  blue  jumper  pin 

located between the transformer and the 9-pin connector. Re-install the base 
shroud. 

OR 

2) 

If the chair has a 2

nd

 generation circuit board, move the red jumper pin located 

between  the  9-pin  connector  and  the  6-pin  connector  labeled  ‘Stand  Safety’ 
from the OFF position to the ON position. Re-install the base shroud. 

 

3.5 

Power Up

 

 

 
Plug  the  chair  power  cord  into  a  wall  receptacle  or  into  the  exam  stand  or  cabinet 
system. The chair control membrane  switches should illuminate to indicate that the 
chair  is  operational.  Perform  a  test  of  all  chair  features  to  verify  proper  function. 
Refer to the operation section of this document for details. 

 
 

4.0 

SYSTEM OPERATION

  

The  OC-2200's  sturdy  design  and  manual  recline  features  make  it  an  excellent  choice  for 
Ophthalmic and Optometric exam rooms. 
 
Each of the chair control membrane switch buttons performs a specific function. Refer to 
Performance and Specifications for the button symbols and their corresponding functions. 

 

4.1 

Elevation Control 

Height adjustment of the manual recline ophthalmic chair model OC-2200 is made by 
pressing the appropriate arrow button on the membrane switch located at the side of 
the back or with an optional footswitch. The chair may also be controlled by the UP, 
DOWN and AUTORETURN buttons on the exam stand. 

 

4.2 

Recline Control 

To recline the patient, press the release button with the thumb and grip the handle in 
the back while standing beside the chair. While the button is depressed move the back 
to  the  desired  position,  then  release  the  button. To  return  the  patient to  the  upright 
position, press the release button again while holding the chair back. The pneumatic 
cylinder will aid in returning the back to the upright position. 

 

4.3 

Swivel Brake Operation

 

To  rotate  the  chair,  move  the  swivel  brake  handle  to  the  released  position 
(perpendicular  to  the  floor),  rotate  the  chair  to  the  desired  position,  then  move  the 
swivel brake handle to the activated position (parallel to the floor) to secure the chair 
in position. 

 

4.4 

Armrest Retraction

 

One or both of the armrests may be pivoted up out of the way if desired. Press the 
latch  release  button  in  the  middle  underside  of  the  armrest  then  lift  up  on  the 

WARNING: Check all power cords and electrical connections for damage prior to

powering up the system.

!

Содержание OC-2200

Страница 1: ...i...

Страница 2: ...ient in a safe condition if a malfunction occurs After Operation Never disconnect any plugs by pulling on the cable Doing so may cause damage to the internal connections If the chair has been idle for...

Страница 3: ...ctions 3 3 5 Power Up 4 4 0 OPERATION 4 4 1 Elevation Control 4 4 2 Recline Control 4 4 3 Swivel Brake Operation 4 4 4 Armrest Retraction 4 4 5 Headrest Position 5 4 6 Footrest Retraction 5 5 0 ADJUST...

Страница 4: ...rest Height 3 5 89mm 1 5 38mm Full Recline Length 75 1905mm Mode of Operation Continuous Use Type of Protection Class 1 protection Against Electrical Shock Ingress of Water Protection Ordinary Protect...

Страница 5: ...es F 23 to 54 Degrees C Relative Humidity Range 30 to 95 Atmospheric Pressure Range 10 1 to 15 4 psi 700 to 1060 hPa Acceptable Storage Time in operating environment 5 years in extreme environment 20...

Страница 6: ...appropriate wall outlet or into the ophthalmic stand 2 Connect the chair control cable to the ophthalmic stand and check the vertical movement of the ophthalmic chair 3 4 Alternate Connections When th...

Страница 7: ...t of the manual recline ophthalmic chair model OC 2200 is made by pressing the appropriate arrow button on the membrane switch located at the side of the back or with an optional footswitch The chair...

Страница 8: ...ng tension Clockwise for more tension counter clockwise for less tension Adjust the clamp tension so that the clamp locking handle closes easily and the pivots at each end of the clamp are secure Repe...

Страница 9: ...snap ring off which holds the release button into the membrane switch Remove the membrane switch from the back casting Next remove the switch blank from the opposite side of the back casting and inst...

Страница 10: ...ounter clockwise and simultaneously pressing down easily replace the fuses Carefully remove the fuse and check for breaks in the filament or for any corrosion at the ends If in doubt test the fuses wi...

Отзывы: