![Topcon BM-7A Скачать руководство пользователя страница 86](http://html1.mh-extra.com/html/topcon/bm-7a/bm-7a_instruction-manual_1128612086.webp)
Appendix
84
Chromaticity
For 2°
1 to 5cd/㎡
: Chromaticity xy 0.005 or less (SLOW mode, AUTO range, standard
sources A)
5cd/㎡ or more
: Chromaticity xy 0.002 or less (SLOW mode, AUTO range, standard
sources A)
Measurement
Angle
Range and Specification
2°
1 to 5cd/㎡ : Chromaticity xy 0.005 or less
5cd/㎡ or more : Chromaticity xy 0.002 or less
1°
4 to 20cd/㎡ : Chromaticity xy 0.005 or less
20cd/㎡ or more : Chromaticity xy 0.002 or less
0.2°
100 to 500cd/㎡ : Chromaticity xy 0.005 or less
500cd/㎡ or more : Chromaticity xy 0.002 or less
0.1°
400 to 2000cd/㎡ : Chromaticity xy 0.005 or less
2000cd/㎡ or more : Chromaticity xy 0.002 or less
Temperature Characteristics
±3% or less (23℃ is regarded as standard in the range of 0℃ to 40℃)
Humidity Characteristics
±3% or less (60%RH is regarded as standard in the range of 60%RH or less)
Calibration Reference
Topcon calibration reference (Standard sources A, 23℃±3℃)
User calibration reference
Switchable
Display
Dot-matrix LCD 20 characters x 4 rows with backlight function
Interface
USB/RS-232C
Analog Output (Function only for analog output model)
Analog voltage output proportion to X
2
, Y, and Z
Power Supply
AC adapter (standard supplied part)
Power Consumption
Approximately 2.5VA when using an AC adapter
Содержание BM-7A