![Topcon BM-7A Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/topcon/bm-7a/bm-7a_instruction-manual_1128612050.webp)
3. Various Settings
48
3.6.2 Inputting the Correction Factor
The correction factor for No. 01 is displayed.
The procedure for inputting the correction factor is as described below.
1
Press the [ROTATION] button to display the data for the number of the correction factor
to be input
[ROTATION]
: The correction factor for the next number is displayed.
[SHIFT]
: Each time the button is pressed, the display changes as shown below.
Reference data → Sample data → Correction factor data → …
[ENTER]
: Returns to the previous screen.
2
Press the [CHANGE] button.
The following screen is displayed.
3
Press the [ROTATION] button, and select the process to be performed (indicated by "*").
DIRECT
: Selected when inputting the correction factor directly.
☞
「Direct Input」
REF. & MEASURE
: Selected when inputting already known luminance and chromaticity data,
performing sample measurement and calculating the correction factor.
☞
「Calculation from Measurement」
BACK
: Selected when returning to the previous screen.
4
Press the [ENTER] button.
FACTOR INPUT
DIRECT
REF. & MEASURE
*BACK
K01 KX= 1.000E+00
SET KY= 1.000E+00
KZ= 1.000E+00
R=NEXT S=PAGE
Содержание BM-7A
Страница 1: ...INSTRUCTION MANUAL LUMINANCE COLORIMETER ...
Страница 2: ......
Страница 21: ...1 Before Use 19 3 Plug in the AC adapter plug into the power outlet ...
Страница 94: ......