31
Reference
Electromagnetic Compatibility
This product conforms to the EMC standard (IEC 60601-1-2 : 2001).
a) MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information provided in the
ACCOMPANYING DOCUMENTS.
b) Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
c) The use of ACCESSORIES, transducers and cables other than those specified, with the
exception of transducers and cables sold by the manufacturer of the EQUIPMENT or SYS-
TEM as replacement parts for internal components, may result in increased EMISSIONS or
decreased IMMUNITY of the EQUIPMENT or SYSTEM.
d) The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equip-
ment. If adjacent or stacked use is necessary, the EQUIPMENT or SYSTEM should be
observed to verify normal operation in the configuration in which it will be used.
e) The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other
than those specified may result in increased EMISSION or decreased IMMUNITY of the
EQUIPMENT or SYSTEM.
Item
Article code
Model No.
Length (m)
RS-232C CROSSING CABLE (shielded)
---
---
4.8
Guidance and manufacturer's declaration - electromagnetic emissions
The ACP-8 is intended for use in the electromagnetic environment specified below.
The customer or the user of the ACP-8 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The ACP-8 uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The ACP-8 is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC61000-3-2
Class B
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies