dBm
%
Comp
Hold
Menu
Local
2nd
DC V
AC V
Ω
Freq
AC+DC
Rel
Rate
Esc
Shift
Trig
Auto
mA
4 Digit Multimeter
TH1941
1
2
POWER
ON
OFF
AC VOLTAGE
SOURCE
MED
AUTO
Period Measurement
Input Impedance =1.1
MΩ
in parallel with 100pF
CAUTION: Maximum Input = 750V RMS or 1000V peak, 3
×
10
7
V-Hz
Figure 3-4
Frequency and Period Measurements
3.6 Measuring Continuity
Under continuity test, when 200
Ω
range is selected, the multimeter alerts you with a beep when a
reading is below a threshold resistance level of 10
Ω
. For other ranges, the beeper will sound if a reading
is below a threshold resistance level listed in table 3-1.
Table 3-1 Threshold resistance levels in Continuity Test
Measuring range
Beeper on
200.00Ω
<10Ω
2.0000KΩ
<100Ω
20.000KΩ
<1kΩ
200.00KΩ
<10kΩ
2.0000MΩ
<100kΩ
20.000MΩ
<1MΩ
3.6.1 Connections
Assuming Model TH1941 is under factory default conditions, the basic procedure is as follows:
1. Connect test leads to
and COM terminals.
2. Select Continuity measurement function by pressing
→
.
3. Connect test leads to the resistance under test as shown in Figure 3-5.
4. Take a reading from the display