8. SYSTEM PARAMETER [
c10
]
1)
Selecting Maintenance code [
C 10
]
2)
Press the START key
■
[
P 0 0
]
External temperature sensor for articles to be sterilized setting
a) Press the time setting
▲ ▼ keys to set the 1 or 2 as equipped external sensor for
articles to be sterilized.
b) Press the STOP key to return to the control panel display as shown below.
Press the STOP key again to end the procedures.
■
[
P 3 5
] Standard Temperature sensor calibration
a) Press the temperature setting
▲ ▼ keys to select the system parameter
[
P 3 5
].
b) Press the START key
c) Press the time setting
▲ ▼ keys to set the calibration value for the temperature
sensor inside the chamber.
C. 1 0
- - -
- -l
P. 0 0
0
- -l
0: not equipped
1: equipped > Monitor timer count down
2: equipped > Monitor timer count down
Temperature to unlock door interlock,
and temperature to complete sterilization.
P. 3 5
0
- -l
P. 3 5
0
- -l
The display differs with the type of the product.