Tomy SX-300 Скачать руководство пользователя страница 5

DME11D327-06

   

 

 

- 4 -

 

6. Error History [

 c.04 

]

 

1) Selecting Maintenance code [ 

C 04

 ] 

 
 
 

2) Press the START key   
 
 
3)Temperature 

▲    ▼  keys : Indicate  last 10 Error Code

 

 

7. number of Operation [

 c.05 

]

 

1) Selecting Maintenance code [ 

C 05

 ] 

 
 
 

2) Press the START key   
 
 
3)Temperature 

▲    ▼  keys : Indicate  number of operation for each mode

 

 

    n.00 : number of total operation 
    n.01 : number of “ sterilize liquid “ operation     
    n.02 : number of “ sterilize normal “ operation 
    n.03 : number of “ sterilize – warming “ operation   
    n.04 : number of “ Heat – warming “ operation 
 
 

7. Door lock monitoring [

 c.08 

]

 

 

1) Selecting Maintenance code [ 

C 08

 ] 

 
 
 

2) Press the START key   

 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

C. 0 4

- - -

- -l

Er. 0

E 0 1

- -l

C. 0 5

- - -

- -l

N.0 0

0 0 0

0 1

C. 0 8

- - -

- -l

L o n

0 0 -

- -l

Lon : Lock the door 
Lof : unlock the door   

Left : Door open /close microswitch 

 

[

o

]

 

=on

   

[

-

]=off

 

Center : Hook microswitch 

[

o

]

 

=on

   

[

-

]=off

 

Right : interlock microswitch 

[

o

] =on    [

-

]=off 

Содержание SX-300

Страница 1: ...toclave SX 300 500 700 SX 300E 500E 700E TOMY SEIKO CO LTD TOMY DIGITAL BIOLOGY CO LTD 3 14 17 Tagara Nerima ku Tokyo 179 0073 Japan e mail info digital biology co jp URL http www digital biology co jp phone 81 3 5971 8160 fax 81 3 3970 6036 ...

Страница 2: ...activate the function of the maintenance code selected When you press the STOP key or CHECK key return to the normal mode 2 System Initialization c 00 1 Press the START key 2 Press the HR TIMER WARM setting or keys to select the SX Model code 1 SX 300 300E 2 SX 500 500E 3 SX 700 700E e g this is case of SX 700 3 Press the START key keep pressing over 3 seconds Temperature display Time display Warm...

Страница 3: ... dummy resistor 123 24 ohm to CN13 on Main Control PCB 5 Press the START key while more than 2 seconds It is displayed as shown below r 2 6 Connect dummy resistor 151 71 ohm to CN13 on Main Control PCB 7 Press the START key while more than 2 seconds It is displayed as shown first of the display 8 Press the STOP key or the CHECK key return to the normal mode Temperature display adjustment is update...

Страница 4: ...s updated 5 Exhaust valve c 03 1 Selecting Maintenance code C 03 2 Press the START key Exhaust valve can be rotated by manually and display about Pulse position and sensor condition a Temperature keys Exhaust valve can be rotated during press the key b START key rotate to original point c Time keys rotate to close position d Time keys rotate to full open position e Exhaust keys Close position leve...

Страница 5: ... number of total operation n 01 number of sterilize liquid operation n 02 number of sterilize normal operation n 03 number of sterilize warming operation n 04 number of Heat warming operation 7 Door lock monitoring c 08 1 Selecting Maintenance code C 08 2 Press the START key C 0 4 l Er 0 E 0 1 l C 0 5 l N 0 0 0 0 0 0 1 C 0 8 l L o n 0 0 l Lon Lock the door Lof unlock the door Left Door open close ...

Страница 6: ...e STOP key again to end the procedures P 3 5 Standard Temperature sensor calibration a Press the temperature setting keys to select the system parameter P 3 5 b Press the START key c Press the time setting keys to set the calibration value for the temperature sensor inside the chamber C 1 0 l P 0 0 0 l 0 not equipped 1 equipped Monitor timer count down 2 equipped Monitor timer count down Temperatu...

Страница 7: ...l panel or the set temperature Enter 11 by pressing the time setting keys d Press the START key to determine the calibration value and store it e Press the STOP key to return to the control panel display as shown below Press the STOP key again to end the procedures P 3 5 8 l Calibration values for the temperature sensor inside the chamber 30 30 Calibration value 1 0 1 P 3 5 1 1 l Calibration value...

Страница 8: ...t temperature Enter 8 by pressing the time setting keys e g If the temperature measured inside the chamber is 1 1 higher than the temperature displayed on the control panel or the set temperature Enter 11 by pressing the time setting keys d Press the START key to determine the calibration value and store it P 3 6 0 l P 3 6 0 l The display differs with the type of the product P 3 6 8 l Calibration ...

Страница 9: ...perature setting keys to select the system parameter P 6 9 b Press the START key c Press the time setting keys to set the altitude segment d Press the STOP key to return to the control panel display as shown below Press the STOP key again to end the procedures C 1 0 l P 6 9 0 l P 6 9 0 l Altitude segment 0 0 1 000m 1 1 000 2 000m 2 2 000 3 000m 89 9 79 5kPa abs P 6 9 1 l C 1 0 l ...

Страница 10: ...TOP key again to end the procedures 9 External Temperature sensor display adjustment c 12 1 Selecting Maintenance code C 12 2 Press the START key e g 23 42 degree C 3 Press the START key while more than 2 seconds It is displayed as shown below r 1 r 1 0 C 1 2 l 244 23 42 Digital data 0 1023 Temperature 0 01 degree C unit Temperature 1 degree C unit 60 P 9 4 0 l P 9 4 1 l Model Code 1 SX 300 300E 2...

Страница 11: ...hown first of the display 8 Press the STOP key or the CHECK key return to the normal mode Temperature display adjustment is updated 10 number of lid open close operation c 22 1 Selecting Maintenance code C 22 2 Press the START key e g 1 234 times open to close Lid open to close 1 time a TIME keys Times of open close can be changed b Press the START key while more than 2 seconds When the buzzer sou...

Отзывы: