High and Low Alarms –
Using Barcodes to set
alarm limits
To generate your alarm barcode you must use the following code template-
*--H____L____*
“H” stands for the high alarm limit and “L” stands for the low alarm limit. Each
of the 4 digit spaces after the defining limits are laid out in this order-
Hundreds, Tens, Units, Decimal
For example,
60°C High Alarm Limit and 30°C Low Alarm Limit will be entered as
*--H0600L0300*
250°C High Alarm Limit and 100°C Low Alarm Limit will be entered as
*--H2500L1000*
5°C High Alarm Limit and 1.2°C Low alarm Limit will be entered as
*--H0050L0012*
You can generate barcodes on your PC and print or use the TME Barcode
Printer.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.