EF200R-C

12

Upper range value

The measuring signal amplitude must be below a certain limit value to ensure that the signals can be
evaluated without error. This results in a maximum permitted flow rate Q

AmpMax

:

Q

[m /h] =

AmpMax

³

350 [m/s] · · D [m]

i

²

π

· 3600 [s/h]

4 ·

³]

m

/

g

k

[

ρ

³]

m

/

g

k

[

 

1

Q

[ /min] =

AmpMax

³

1148 [ /s] · · D [ft]

i

²

π

· 60 [s/min]

4 ·

0.0624 [lbm/ ]

³

  A0034316

Q

AmpMax

Maximum measurable flow rate based on signal amplitude

D

i

Internal diameter of measuring tube (corresponds to dimension K→  See page 41)

ρ

Density

For gas applications, an additional restriction applies to the upper range value with regard to the
Mach number in the measuring device, which must be less than 0.3. The Mach number Ma describes
the ratio of the flow velocity v to the sound velocity c in the fluid.

Ma =

v [m/s]
c [m/s]

Ma =

  A0034321

Ma

Mach number

v

Flow velocity

c

Sound velocity

The corresponding flow rate can be derived using the nominal diameter.

Q

[m /h] =

Ma = 0.3

³

0.3 · c [m/s] · · D [m]

i

²

π

4

· 3600 [s/h]

Q

[ /min] =

Ma = 0.3

³

0.3 · c [ft/s] · · D [ft]

i

²

π

4

· 60 [s/min]

  A0034337

Q

Ma = 0.3

Restricted upper range value is dependent on Mach number

c

Sound velocity

D

i

Internal diameter of measuring tube (corresponds to dimension K→  See page 41)

ρ

Density

Содержание EF200R-C

Страница 1: ...172 65760M 01 EF200R C Vortex owmeter 7 June 2022 EF200R C Vortex owmeter Copyright 2022 by TLV CO LTD All rights reserved...

Страница 2: ...Inlet and outlet runs Length of connecting cable Mounting the transmitter housing Environment Ambient temperature range Storage temperature Climate class Degree of protection Vibration and shock resi...

Страница 3: ...de the device Inner ground terminal Connects the protectiv earth to the mains supply Outer ground terminal Connects the device to the plant grounding system Symbols for certain types of information Sy...

Страница 4: ...endent of the ow velocity and the uid properties viscosity and density Independent of the type of substance under measurement steam gas or liquid The primary measuring signal is linear to the ow After...

Страница 5: ...e temperature measurement Lifelong calibration Experience has shown that recalibrated measuring devices demonstrate a very high degree of stability compared to their original calibration The recalibra...

Страница 6: ...tion by using the Prowirl with integrated line size reduction 1 Reducer element 2 Straight pipe segment as the inlet run min 15 DN upstream from the vortex meter 3 Straight pipe segment as the outlet...

Страница 7: ...via four line Via operating tools A0013471 Sensor EF200R C A0034075 Pressure measuring cell 1 A0034080 1 Pressure measuring cell Versions Pressure components Pressure measuring cell 40 bar a Material...

Страница 8: ...ange is dependent on the nominal diameter the uid and environmental in uences The following speci ed values are the largest possible ow measuring ranges Qmin to Qmax for each nominal diameter Dependin...

Страница 9: ...6 to 1 100 6R 3 to 140 15 to 1900 8R 6 8 to 320 34 to 4300 Flow velocity D i v Q A0033468 Di Internal diameter of measuring tube corresponds to dimension K See page 41 v Velocity in measuring tube Q F...

Страница 10: ...uid and the nominal diameter is used to calculate the corresponding ow rate Q m h Re 5000 5000 Pa D m s i 3600 s h 4 kg m Q h Re 5000 5000 lbf D ft s ft i 60 s min 4 lbm A0034302 QRe 5000 Flow rate i...

Страница 11: ...n D ft i 60 s min 4 0 0624 lbm A0034304 QAmpMin Minimum measurable ow rate based on signal amplitude vAmpMin Minimum measurable ow velocity based on signal amplitude Di Internal diameter of measuring...

Страница 12: ...y For gas applications an additional restriction applies to the upper range value with regard to the Mach number in the measuring device which must be less than 0 3 The Mach number Ma describes the ra...

Страница 13: ...tput 1 4 20 mA passive Resolution 1 A Damping Adjustable 0 0 to 999 9 s Assignable measured variables Volume ow Corrected volume ow Mass ow Flow velocity Temperature Pressure Calculated saturated stea...

Страница 14: ...0 to 1000 Hz Damping Adjustable 0 to 999 s Pulse pause ratio 1 1 Assignable measured variables Volume ow Corrected volume ow Mass ow Flow velocity Temperature Calculated saturated steam pressure Tota...

Страница 15: ...the interface failure information is displayed as follows Current output 4 to 20 mA 4 to 20 mA Failure mode Choose from 4 to 20 mA in accordance with NAMUR recommendation NE 43 4 to 20 mA in accordan...

Страница 16: ...RB including line resistance must be observed to ensure adequate terminal voltage at the device In doing so observe the minimum terminal voltage For U S 17 9 to 18 9 V R B US 17 9 V 0 0036 A For U S 1...

Страница 17: ...1 2 3 4 4 20 mA passive Pulse frequency switch output passive 1 Output 1 must always be used output 2 is optional Connecting cable for remote version Transmitter and sensor connection housing In the c...

Страница 18: ...o the signal cable when lifting the transmitter housing Lift the transmitter housing plug the signal cable out of the connection board of the wall holder and remove the transmitter housing 4 Release t...

Страница 19: ...color Connecting cable 1 Supply voltage Brown 2 Grounding White 3 RS485 Yellow 4 RS485 Green Connecting cable option mass pressure temperature compensated Order code for Sensor version DSC sensor meas...

Страница 20: ...ation DC 1 V Sensor version DSC sensor measuring tube Increase in minimum terminal voltage Mass steam 316L 316L integrated pressure temperature measurement DC 1 V For information about the load See p...

Страница 21: ...tored Electrical connection Connecting the transmitter 1 1 A0033480 1 Cable entries for inputs outputs Remote version connection Connecting cable 3 1 2 A0033481 Connecting cable connection 1 Wall hold...

Страница 22: ...s to comply with EMC requirements observe cable speci cations 4 Analog display unit observe maximum load 5 Transmitter Pulse frequency output 1 2 3 12345 A0028761 Connection example for pulse frequenc...

Страница 23: ...s that apply in the country of installation must be observed The cables must be suitable for the minimum and maximum temperatures to be expected Signal cable Current output 4 to 20 mA A shielded cable...

Страница 24: ...ordered with integrated overvoltage protection for diverse approvals Order code for Accessory mounted option NA Overvoltage protection Input voltage range Values correspond to supply voltage speci ca...

Страница 25: ...AmpMin v m s AmpMin D m i 3600 s h 4 m g k m g k 1 Q min AmpMin v s AmpMin D ft i 60 s min 4 0 0624 lbm A0034304 Remax De ned by internal diameter of measuring tube Mach number and maximum permitted v...

Страница 26: ...Sensor version Mass integrated temperature measurement Mass integrated pressure temperature measurement 1 Process pressure bar abs Flow velocity m s ft s Reynolds number range Measured value deviatio...

Страница 27: ...diameter mismatch correction can compensate for the e ects The di erence between the internal diameter of the ordered process connection and the internal diameter of the mating pipe used must be take...

Страница 28: ...if the measured volume is increased Repeatability is not a device characteristic but a statistical variable that is dependent on the boundary conditions indicated Response time If all the con gurable...

Страница 29: ...n of 16 mA Temperature coe cient at zero point 4 mA 0 02 10 K Temperature coe cient with span 20 mA 0 05 10 K Pulse frequency output o r of reading Temperature coe cient Max 100 ppm o r Installation M...

Страница 30: ...s 200 C 392 F orientation B is not permitted for the wafer version Prowirl D with nominal diameters of DN 100 4 and DN 150 6 3 In the case of hot media e g steam or uid temperature TM 200 C 392 F orie...

Страница 31: ...cable length A L A0019211 A Minimum spacing in all directions L Required cable length The following dimensions must be observed to guarantee problem free access to the device for service purposes A 10...

Страница 32: ...N 4 2 20 DN 5 DN 5 20 DN 5 DN 6 17 DN 8 h5 DN h 7 50 DN 5 DN 9 40 DN 5 DN 8 5 DN DN 25 1 DN 40 1 A0019189 5 DN 5 DN 5 DN 25 DN Minimum inlet and outlet runs with various ow obstructions DN Pipe diamet...

Страница 33: ...nation pipe Double elbow 3D 2 90 elbows opposite not on one lane reducer etc 5 DN 5 DN 5 DN 5 DN 5 DN Flow conditioner If the inlet runs cannot be observed the use of a ow conditioner is recommended T...

Страница 34: ...e remote version observe the maximum permitted cable length Lmax 30 m 90 ft The value for the cable length must be calculated if the cable cross section di ers from the speci cation For detailed infor...

Страница 35: ...cover from TLV Storage temperature All components apart from the display modules 50 to 80 C 58 to 176 F Climate class DIN EN 60068 2 38 test Z AD Degree of protection Transmitter As standard IP66 67 t...

Страница 36: ...ium temperature range Pressure measuring cell 40 bar 580 psi abs 40 to 100 C 40 to 212 F Seals Description Medium temperature range Graphite standard 200 to 400 C 328 to 752 F Pressure temperature rat...

Страница 37: ...i 0 100 200 300 400 500 600 700 800 900 400 200 0 800 200 600 400 PN10 PN16 PN25 PN40 0 10 20 30 40 50 60 200 100 0 100 200 300 400 C F bar psi 0 100 200 300 400 500 600 700 800 900 400 200 0 800 200...

Страница 38: ...200 300 400 500 600 700 800 900 200 0 800 200 600 400 PN40 Class 300 PN40 Class 150 0 10 20 30 40 50 60 200 100 0 100 200 300 400 C F bar psi 0 100 200 300 400 500 600 700 800 900 200 0 800 200 600 40...

Страница 39: ...ge 26 The OPL may only be applied for a limited period of time The MWP maximum working pressure for the sensors depends on the lowest rated element with regard to pressure of the selected components i...

Страница 40: ...of the sensor the device is set at the factory at the very maximum to the OPL value of the process connection If using the entire sensor range select a process connection with a higher OPL value Senso...

Страница 41: ...on respective ange connection DN Reduction to DN A 1 B C1 E G H I K Di L mm mm mm mm mm mm mm mm mm mm mm 25R 15 140 2 51 7 88 5 252 159 9 58 2 101 7 13 9 2 40R 25 140 2 51 7 88 5 258 159 9 58 2 101...

Страница 42: ...s version 1 For high temperature low temperature version values 29 mm 2 Dependent on respective ange connection DN Reduction to DN A B C E G K Di L mm mm mm mm mm mm mm mm mm 25R 15 107 3 60 0 47 3 22...

Страница 43: ...295 8 22 24 146 3 300 Raised face according to DIN EN 1092 1 Form B1 Ra 6 3 to 12 5 m 1 In compliance with ISO 13359 for DN 150 Flange connection dimensions according to DIN EN 1092 1 PN 16 Triple ce...

Страница 44: ...8 22 25 112 0 300 Raised face according to DIN EN 1092 1 Form B1 Ra 6 3 to 12 5 m 1 In compliance with ISO 13359 for DN 100 to 150 Flange connection dimensions according to DIN EN 1092 1 PN 25 Triple...

Страница 45: ...160 8 18 25 0 56 5 200 100R 80 235 190 8 22 26 5 87 0 250 150R 100 300 250 8 26 31 0 112 0 300 200R 150 375 320 12 30 36 5 146 3 300 Raised face according to DIN EN 1092 1 Form B1 Ra 6 3 to 12 5 m 1 I...

Страница 46: ...Ra 6 3 to 12 5 m 1 In compliance with ISO 13359 for DN 15 to 100 Flange connection dimensions according to ASME B16 5 Class 150 Schedule 40 Triple certi ed material 1 4404 F316 F316L DN mm Reduction t...

Страница 47: ...9 4 241 3 8 22 4 26 0 112 0 300 Raised face according to ASME 16 5 Ra 3 2 to 6 3 m 200R 150 381 0 330 2 12 25 4 41 1 146 3 300 Raised face according to ASME 16 5 Ra 3 2 to 6 3 m Flange connection dime...

Страница 48: ...9 6 168 1 8 22 4 28 9 56 5 200 100R 80 254 0 200 2 8 22 4 31 8 87 0 250 150R 100 317 5 269 7 12 22 4 39 0 112 0 300 Raised face according to ASME 16 5 Ra 3 2 to 6 3 m Flange connection dimensions acco...

Страница 49: ...00 Raised face according to JIS 2220 Ra 3 2 to 6 3 m Flange connection dimensions according to JIS B2220 20K Schedule 40 Triple certi ed material 1 4404 F316 F316L DN mm Reduction to DN mm A mm B mm C...

Страница 50: ...316L DN mm Reduction to DN mm A mm B mm C mm D mm E mm L mm 25R 15 125 90 4 19 18 5 22 0 200 40R 25 140 105 4 19 19 0 30 0 200 50R 40 155 120 8 19 22 0 45 0 200 80R 50 200 160 8 23 27 0 56 5 200 100R...

Страница 51: ...50 221 0 D2 20 0 200 274 0 D2 26 3 Used in combination with anges according to DIN EN 1092 1 PN 10 1 4404 316 316L Order code for Accessory enclosed option PF DN mm Centering diameter D11 D2 2 mm s mm...

Страница 52: ...ination with anges according to DIN EN 1092 1 PN 25 1 4404 316 316L DN mm Centering diameter mm D11 D2 2 s mm 15 54 3 D2 2 0 25 74 3 D1 3 5 40 95 3 D1 5 3 50 110 0 D2 6 8 80 145 3 D2 10 1 100 171 3 D1...

Страница 53: ...88 2 D2 5 3 50 106 6 D2 6 8 80 138 4 D1 10 1 100 176 5 D2 13 3 150 223 5 D1 20 0 200 274 0 D2 26 3 250 340 0 D1 33 0 1 The ow conditioner is tted at the outer diameter between the bolts 2 The ow condi...

Страница 54: ...t the outer diameter between the bolts 2 The ow conditioner is tted at the indentations between the bolts Used in combination with anges according to JIS B2220 20K 1 4404 316 316L DN mm Centering diam...

Страница 55: ...ree cleaning is not possible B D E F G L C A0033851 Order code for Sensor version DSC sensor measuring tube Option Mass steam 316L 316L integrated pressure temperature measurement DN mm B mm C mm D mm...

Страница 56: ...e inner diameter line size reduction DN Reduction to DN A 1 B C1 E G H I K Di L in in in in in in in in in in in 1R 5 52 2 04 3 48 9 92 6 3 2 29 4 0 55 2 1 R 1 5 52 2 04 3 48 10 2 6 3 2 29 4 0 96 2 2R...

Страница 57: ...EF200R C 57 Transmitter remote version A B C G Q Q F T A0033796 A 1 B C1 F G Q T in in in in in in in 5 52 2 04 3 48 10 6 3 4 21 7 52 1 For version with overvoltage protection value 0 31 in...

Страница 58: ...ction to DN A B C E 1 G K Di L in in in in in in in in in 1R 4 22 2 36 1 86 8 86 3 72 0 55 2 1 R 1 4 22 2 36 1 86 9 09 3 72 0 96 2 2R 1 4 22 2 36 1 86 9 41 3 72 1 5 2 3R 2 4 22 2 36 1 86 9 65 3 72 1 9...

Страница 59: ...hedule 40 Triple certi ed material 1 4404 F316 F316L DN in Reduction to DN in A in B in C in D in E in L in 1R 4 26 3 12 4 0 62 0 71 0 87 7 87 1 R 1 5 3 88 4 0 62 0 71 1 18 7 87 2R 1 6 4 75 4 0 75 0 7...

Страница 60: ...B16 5 Ra 125 to 250 in Flange connection dimensions according to ASME B16 5 Class 150 Schedule 80 Triple certi ed material 1 4404 F316 F316L DN in Reduction to DN in A in B in C in D in E in L in 1R 4...

Страница 61: ...R 4 11 8 10 6 12 0 88 1 52 4 41 11 80 8R 6 15 13 12 1 1 62 5 76 11 80 Raised face according to ASME B16 5 Ra 125 to 250 in Flange connection dimensions according to ASME B16 5 Class 300 Schedule 80 Tr...

Страница 62: ...The ow conditioner is tted at the outer diameter between the bolts 2 The ow conditioner is tted at the indentations between the bolts Used in combination with anges according to ASME B16 5 Class 300 1...

Страница 63: ...eaning is not possible B D E F G L C A0033851 Order code for Sensor version DSC sensor measuring tube Option DA Mass steam 316L 316L integrated pressure temperature measurement DN in B in C in D in E...

Страница 64: ...re version values 0 2 kg 1 For high temperature low temperature version values 0 4 lbs DN mm Internal diameter mm Weight kg 25R 15 6 1 40R 25 10 1 50R 40 12 1 80R 50 16 1 100R 80 23 1 150R 100 42 1 20...

Страница 65: ...in kg 1 For high temperature low temperature version values 0 2 kg Weight in US units All values weight refer to devices with ASME B16 5 Class 300 Sch 40 anges Weight information in lbs 1 For high tem...

Страница 66: ...40 2 4 150 PN 10 16 PN 25 40 6 3 7 8 200 PN 10 PN 16 25 PN 40 11 5 12 3 15 9 250 PN 10 to 25 PN 40 25 7 27 5 DN1 mm Pressure rating Weight kg 15 Class 150 Class 300 0 03 0 04 25 Class 150 Class 300 0...

Страница 67: ...80 150 10K 20K 4 5 5 5 200 10K 20K 9 2 250 10K 20K 15 8 19 1 DN1 in Pressure rating Weight lbs Class 150 Class 300 0 07 0 09 1 Class 150 Class 300 0 3 1 Class 150 Class 300 0 7 2 Class 150 Class 300 1...

Страница 68: ...G Nickel plated brass Connecting cable for remote version Standard cable PVC cable with copper shield g n i s u o h r e t t i m s n a r T Materials Compact version Aluminum coated compact Aluminum Al...

Страница 69: ...rane Stainless steel 1 4435 316L Non wetted parts Housing Stainless steel 1 4404 Siphon 4 Stainless steel 1 4571 Adjusting nut Stainless steel 1 4571 Pressure gauge valve Stainless steel 1 4571 Welded...

Страница 70: ...easuring tube option Copper Housing support Stainless steel 1 4408 CF3M Screws for DSC sensor Stainless steel A2 80 according to ISO 3506 1 304 Accessories Protective cover Stainless steel 1 4404 316L...

Страница 71: ...event logbook No need to recon gure E cient diagnostics increase measurement availability Troubleshooting measures can be called up via the device and in the operating tools Diverse simulation options...

Страница 72: ...of the Australian Communications and Media Authority ACMA HART certification HART interface The measuring device is certi ed and registered by the FieldComm Group The measuring system meets all the re...

Страница 73: ...tromagnetic compatibility EMC requirements NAMUR NE 21 Electromagnetic compatibility EMC of industrial process and laboratory control equipment NAMUR NE 32 Data retention in the event of a power failu...

Страница 74: ...r the sensor Measuring device Documentation code EF200 C 172 65765M Operating Instructions Measuring device Documentation code EF200R C 172 65757M Description of Device Parameters Measuring device Doc...

Страница 75: ...oper storage maintenance or repair or 7 operation of the Products not in accordance with instructions issued with the Products or with accepted industry practices or 8 use for a purpose or in a manner...

Страница 76: ...OTHER PROPERTY DAMAGE TO BUYER S OR THE FIRST END USER S PRODUCT DAMAGE TO BUYER S OR THE FIRST END USER S PROCESSES LOSS OF USE OR OTHER COMMERCIAL LOSSES WHERE DUE TO OPERATION OF LAW CONSEQUENTIAL...

Страница 77: ...3131 Australia Tel Fax 61 0 3 9873 5610 61 0 3 9873 5010 In East Asia 36 Kaki Bukit Place 02 01 02 Singapore 416214 Tel Fax 65 6747 4600 65 6742 0345 5 F Building 7 No 103 Caobao Road Xuhui District...

Отзывы: