Display Appearance:
The down arrows select the various display functions:
Decrease Total for -ve flow
: The meter will not register negative
flow unless this box is checked.
For reverse flow the rate will show a “-“ sign before the rate and
the total will be reduced accordingly. One of the transistor outputs
(PNP or NPN on Pin 6, 7 or 8) could be configured to give a logic
level when reverse flow is detected.
Pulses are only output on reverse flow, when this box is selected.
Display Mode:
The down arrow selects from the various display
and Metraflow functions.
The options are:
Rate – The display will show flow rate only.
Total – The display will show total flow only.
Both – The display can be cycled from rate to total using the
left hand button on the Meter or the left button in the Flow
Details window on the Interface. A remote input can be
configured to perform this function (See: Pin Assignments).
Rate DP Position
:
Use the drop down menu to choose the required
decimal point position.
Total DP Position:
Use the drop down menu to choose the
required decimal point position.
If the decimal point is set too low and the display value exceeds
the setting capacity the display will show “oF”