B A T T E R Y R E P L A C E M E N T
Battery Type
Renata CR2325
Replacement Interval
Annually*
*If logging intervals of less than five seconds are used continuously, the battery life of
the unit will be reduced and the battery will need to be replaced more frequently.
Before replacing the battery the data logger must be stopped.
When replacing the battery, wait at least one minute after removing the old battery
before fitting the new one.
Data stored on the logger will be retained after a battery is replaced.
If used at low temperatures the data logger should be allowed to warm to room
temperature before it is opened to avoid condensation forming inside the unit.
A service kit for the Tinytag Transit 2, containing a replacement battery
and servicing instructions, can be ordered from your supplier using the
part number SER-9514.