22
MD100_4 10/2014
GB
Technical Data
Technical Data
Instrument
single
wavelength, direct reading colorimeter
Light source:
LED,
interference filters (IF) and photosensor in transparent
cell chamber. Wavelength specifications of the IF:
660 nm
= 5 nm
Wavelength accuracy
± 1 nm
Photometric accuracy*
3% FS (T = 20° C – 25° C)
Photometric resolution
0.01 A
Power supply
4 batteries (AAA/LR 03)
Operating time
17hr operating time or 5000 test measurements in
continuous mode when display backlight is off
Auto-OFF
automatic switch off
10 minutes after last keypress
Display
backlit LCD (on keypress)
Storage
internal ring memory for 16 data sets
Interface
IR interface for data transfer
Time
real time clock und date
Calibration
user and factory calibration
resetting to factory calibration possible
Dimensions
155 x 75 x 35 mm (LxWxH)
Weight
approx. 260 g (incl. batteries)
Ambient conditions
temperature: 5– 40°C
rel. humidity: 30–90% (non-condensing)
Waterproof
floating; as defined in IP 68 (1 hour at 0.1 meter)
CE
Certificate for Declaration of CE-Conformity
at www.lovibond.com
*measured with standard solutions
To ensure maximum accuracy of test results, always use the reagent systems supplied by