TYPE 5867
PORTABLE ‘Q’ METER

OPERATING INSTRUCTIONS

5867 Operating Instructions : Issue 3

www.tinsley.co.uk

8

REMOTE DIN (Front panel view) 

RS232 commands

Function

Remarks

?

63dec

Help menu

Send this list

K

75dec

Get for model version & board issues

S

83dec

Get internal status

Send 6 bytes

D

68dec

Get current instrument display value

Range+Gain+reading (6 bytes)

R

82dec

Read next value or force read

Range+Gain+reading (6 bytes)

Gxn

71dec

Set Range & Gain where x=A/B/C/D/E & n=0 to 15.

Responds with xn

T

84dec

Get Internal Temperature

‘TempCo = xC’

C

67dec

Do Calibration cycle

A

65dec

Get Internal Voltages

I

73dec

Get Unit Itent string

‘Tinsley,5867,xVx’

L{OPT}

Start logging readings QPT –Output Format

Q,xxx,P,xxx,T,xxx where

Q=value, P = Reading interval, T = temp

value is present (xxx) if QPT 
are specified.

P{x,y}

Set pulse duration charge, read

** Engineering mode only

E{x}

Eeprom diagnostics where x is

** Engineering mode only

P – dump to serial channel
D – Reset to default
C = Clear

W{x}

Run diagnostics function

** Engineering mode only

F – Flash all LEDS
P – Set processor port to x
1 – Strobe on
0 – Strobe off
Q – Power off
Rnx – relay n on(1) off(0) i.e. R20 – Relay 2 off.
Ax – Get ADC channel x value

Содержание 5867

Страница 1: ...S TINSLEY A DIVISION OF HARTEST PRECISION INSTRUMENTS LTD SUBJECT TO CHANGE WITHOUT NOTICE This manual superseded all previous versions please keep for future reference IMPORTANT Please read these ins...

Страница 2: ...features warning lights which are illuminated when appropriate Accessories supplied as standard 1 mains cord 1 operating instructions English 2 test leads colour coded length 3m each supplied with sel...

Страница 3: ...The use of makeshift fuses and the short circuiting of fuse holders is prohibited Whenever it is likely that the protection has been impaired the apparatus shall be made inoperative and be secured aga...

Страница 4: ...s built in and the instrument may be connected to a voltage supply of between 90 to 260V AC The instrument may be operated with a flat battery provided that the unit is powered to the mains in this wa...

Страница 5: ...temperature of the unit The display will then show which capacitance range and measuring range was last used and this will be shown on the display until it is next changed by the operator The last use...

Страница 6: ...dicating that a higher range should be selected which can be achieved by pressing the up arrow key A suitable range as would be chosen by the Auto range method described above would be to pick a range...

Страница 7: ...titled Oven This will flash when the oven is within 5 C of the standard 40 C REMOTE This illuminates when the Q meter is being controlled remotely and is lit whilst commands are being transferred via...

Страница 8: ...nds with xn T 84dec Get Internal Temperature TempCo xC C 67dec Do Calibration cycle A 65dec Get Internal Voltages I 73dec Get Unit Itent string Tinsley 5867 xVx L OPT Start logging readings QPT Output...

Страница 9: ...ices at any time without prior notice For further details please refer to our website www tinsley co uk 1 Warner Drive Springwood Industrial Estate Braintree Essex CM7 2YW Tel 01376 335271 E mail info...

Отзывы: