TYPE 5867
PORTABLE ‘Q’ METER

OPERATING INSTRUCTIONS

5867 Operating Instructions : Issue 3

www.tinsley.co.uk

3

WARNING

When the apparatus is connected to its mains supply and the covers are opened, it is likely to expose live parts.

The apparatus must be disconnected from all voltage sources before it is opened for any adjustment, replacement,
maintenance or repair.

Any adjustment, maintenance or repair of the opened apparatus under voltage shall be avoided as far as possible, and if
inevitable, shall be carried out only by a skilled person who is aware of the hazard involved.

Make sure that only fuses with the required current rating and of the specified type are used for replacement. The use of
makeshift fuses, and the short circuiting of fuse holders is prohibited.

Whenever it is likely that the protection has been impaired, the apparatus shall be made inoperative and be secured
against any unintended operation and returned to our factory or agent for rectification.

3. CASE DESIGN

The case is ruggedly constructed from an ABS/polycarbonate alloy. A strong internal sub frame ensures that the 5867 will
withstand the harshest of environments. The front panel is a reverse printed polycarbonate overlay with clear and
unambiguous text.

4. SPECIFICATION

Digital display

8 digit, LED 0.8” height
Range indicator, 999 count and status indicators via 3 LED’s

Working temperature

0…+40°C 
Relative Humidity max. 80%

Normal temperature

20°C

Oven temperature

40°C

Storage temperature

-20°C…+50°C

Mains connection

90 - 260V

Size (mm)

24 x 28 x 12cm

Weight

4.5 kg

Measuring frequency

250ms

Содержание 5867

Страница 1: ...S TINSLEY A DIVISION OF HARTEST PRECISION INSTRUMENTS LTD SUBJECT TO CHANGE WITHOUT NOTICE This manual superseded all previous versions please keep for future reference IMPORTANT Please read these ins...

Страница 2: ...features warning lights which are illuminated when appropriate Accessories supplied as standard 1 mains cord 1 operating instructions English 2 test leads colour coded length 3m each supplied with sel...

Страница 3: ...The use of makeshift fuses and the short circuiting of fuse holders is prohibited Whenever it is likely that the protection has been impaired the apparatus shall be made inoperative and be secured aga...

Страница 4: ...s built in and the instrument may be connected to a voltage supply of between 90 to 260V AC The instrument may be operated with a flat battery provided that the unit is powered to the mains in this wa...

Страница 5: ...temperature of the unit The display will then show which capacitance range and measuring range was last used and this will be shown on the display until it is next changed by the operator The last use...

Страница 6: ...dicating that a higher range should be selected which can be achieved by pressing the up arrow key A suitable range as would be chosen by the Auto range method described above would be to pick a range...

Страница 7: ...titled Oven This will flash when the oven is within 5 C of the standard 40 C REMOTE This illuminates when the Q meter is being controlled remotely and is lit whilst commands are being transferred via...

Страница 8: ...nds with xn T 84dec Get Internal Temperature TempCo xC C 67dec Do Calibration cycle A 65dec Get Internal Voltages I 73dec Get Unit Itent string Tinsley 5867 xVx L OPT Start logging readings QPT Output...

Страница 9: ...ices at any time without prior notice For further details please refer to our website www tinsley co uk 1 Warner Drive Springwood Industrial Estate Braintree Essex CM7 2YW Tel 01376 335271 E mail info...

Отзывы: