- 4 -

M

ODEL 

CI-50 

NOTE: 

The  CI-50  is  designed  to  be  used  by  trained  individuals.  Many 

jurisdictions have laws or recommendations for the type of training necessary in 
order  to  handle  high  voltage  connections.  Consult  your  company  or  jurisdiction 
for guidance in the type of training necessary to use this instrument properly.

 

Operation 

Once  connections  are  made  to  the  circuit  to  be  interrupted,  the  instrument  is  ready  for 
operation.  Turn  the  instrument  Power  toggle  switch  to  On  (Up).  Verify  programming  by  the 
pressing View Program button. If changed to programming are necessary, turn the Display off 
(if on) by again pressing the View Program button. Press the Program button and proceed as 
described in the 

Programming 

section above. 

The  instrument  begins  operation  once  the  Start  /  Stop  button  is  pressed.  To  end  operation, 
press the Start / Start button once more, or turn the Power toggle switch to Off (Down). 

The Model CI-50 is a normally Open (Off) instrument. This means that when not in operation or 
when the power is turned off, the circuit through the instrument is Open (Off). When a program 
begins, the program always starts in the Open (Off) position. 

Batteries 

The Model CI-

50 is battery operated. The instrument uses four (4) “AA” size alkaline batteries. 

It is not recommended that rechargeable batteries be used in this instrument. 
The CI-50 has a Low Battery indication. The Low Batt LED to the Right of the 
Power  toggle  switch  will  begin  to  flash  when  a  low  battery  condition  is 
detected.  When  the  Low  Batt  LED  is  flashing,  it  is  recommended  that  the 
batteries be replaced.  

Battery  life  for  the  Model  CI-50  is  expected  to  be 
approximately  50  hours.  Program  timing  and  ambient 
temperature will have an effect on the battery life. 

When  replacing  the  battery,  observe  polarity.   

The  “AA” 

batteries are placed with two (2) positive (+) up and two (2) 
positive  (+)  down.  Reversed  polarity  will  cause  the 
instrument  to  be  non-functional.    The  CI-50  will  not  be 
harmed by such a connection, but it will not operate. 
It is useful to carry  extra batteries when working in remote 
field locations. 

CORROSION MITIGATION INSTRUMENTATION 

2828 FM 758, NEW BRAUNFELS, TX 78130 TEL: (830) 253-5621 

Содержание CI-50

Страница 1: ...and duration of the spike the solid state circuitry of the CI 50 and most other current interrupters made today can become significantly damaged It is recommended that an oscilloscope be connected to...

Страница 2: ...on for guidance in the type of training necessary to use this instrument properly Programming the CI 50 The Model CI 50 is capable of interrupt cycles of 0 1 sec Closed Seconds On and Open Seconds Off...

Страница 3: ...s be turned Off Power toggle switch to Off down when making or breaking connections Connecting the Model CI 50 is done using the supplied cable A single plug in type connector on the end of the cable...

Страница 4: ...l CI 50 is a normally Open Off instrument This means that when not in operation or when the power is turned off the circuit through the instrument is Open Off When a program begins the program always...

Страница 5: ...he instrument cannot be programmed The Program button will not work until the Display is turned off by again pressing the View Program button or by automatically shutting off Once the Display is off t...

Страница 6: ...en Cycles 0 1 sec to 99 9 sec Programmable Battery Size AA size Battery Type Alkaline Battery Volts 4 1 5v AA Cable Size 10 AWG 6 0 mm2 Cable Type Stranded copper PVC jacketed Black Cable Length 6 1 8...

Страница 7: ...This step becomes especially important when the DC output of the rectifier is 50 or more of the maximum current the interrupter can handle In the case of the CI 50 which is a 50 Amp max instrument a...

Отзывы: