11. TROUBLESHOOTING
Preface
All maintenance operations must be done by skilled personnel!
This chapter lists some of the most probable problems that could affect proper the
wall stretcher operation. Solutions to these problems have been offered in the
troubeshooting guide.
If the problems persist and repeat themselves with some frequency even after
applying the remedies suggested in this chapter, you are kindly asked to contact
our after – sales department.
ERROR
CAUSE
CORRECTION
The head part will not go up
when you press the lever
1.
Excessive load on the
backrest part
2.
Damage of gas springs
3.
Damage of trigger
1.
Reduce the load and try
again by raising
2.
Call a technician to replace
the gas spring
3.
Call a technician to replace
the trigger
The mattress platform of the
shower trolley is not possible to
be raised
1.
It is possible that the battery
is completely flat or
defective or is not properly
connected
2.
It is possible that the hand
control panel is defective
3.
It is possible that the actuator
is defective
4.
It is possible that the feeding
cable is defective
5.
It is possible that the control
box is defective
1. Connect battery to a charger
or call a technician to replace
the battery
2.Call a technician to replace the
hand control panel
3.Call a technician to replace the
actuator
4.Check if all the electric cables
are connected correctly, or call
a technician to replace the
feeding cable
5.Call a technician to replace the
control box
12. SCRAPPING AND DISPOSAL
The Electric wall mounted shower stretcher art. 155040 and Electric wall mounted changing table art.
155140, manufactured by TIMO d.o.o. Ajdovš
č
ina, is constructed mainly from ferrous and non-ferrous
metals, thermoplastic materials.
In the case of demolition and disposal, parts considered as special waste must be processed in
accordance with curent law.