4033A Operations and Maintenance Manual
15
Example: S0000,0001,2
This example shows that the system has no current alarms, output 1 alarmed since last status
request, and the input type is DC IRIG. With no input signal, all status bits will report 0.
3.2.2 Checking Model Number and Software Version
Both commands are case sensitive. You do not need to type a carriage return or line feed.
To check the TSC model number:
Type:
I
The system returns
I4033A-00\r\n
.
To check the software version
Type:
V
The system returns
V
xx
\r\n
where
xx
is the software version.
Example: V80.
Sending an
I
command will turn on the three front panel input LEDs for 2 seconds before
returning to normal operation. This provides the capability for the user to ensure they are
communicating over the Ethernet connection to the correct device via front panel feedback.
3.3 Understanding Alarm Output
The 4033A automatically publishes alarms to the CNR port as they occur.
The alarms appear in the format
ALARM
cccc
where
cccc
is the summary status of the input and
output signals. The format of the alarm status is identical to the “S” command response
cccc
for
current status.
Example: ALARM0008
This example shows that output 4 failed.
Содержание TSC 4033A
Страница 4: ...This page intentionally left blank...
Страница 17: ...4033A Operations and Maintenance Manual 17 This page intentionally left blank...
Страница 19: ...4033A Operations and Maintenance Manual 19 This page intentionally left blank...
Страница 21: ...4033A Operations and Maintenance Manual 21 This page intentionally left blank...
Страница 26: ......
Страница 27: ...4033A Operations and Maintenance Manual 1...