Switch at the Power Entry Module or the power supply cord set is not accessible when the 
unit is mounted inside an enclosed rack.  

 

After power-on, the unit runs a self-test in the following sequence: 

 

All LEDs will be turned on GREEN 

 

Each output LED will sequentially be turned on RED and then GREEN. 

 

The input and output LEDs will be in the appropriate state relative to their respective 
signal status.  

2.4.3.2 Input Signal 

Connect the input signal cable to the 

INPUT 

BNC connector on the back of the unit. 

2.4.3.3 Output Signals 

Connect up to fifteen cables to the 

OUTPUT

 BNC connectors on the rear panel of the unit to 

supply users with a copy of the input signal.  

2.4.3.4 LAN - Ethernet Network Port  

Connect a network LAN cable (not supplied with the unit) to the LAN port on the back of the 
unit. 

2.5 Assigning a Static IP Address 

 

The 4033A contains a Lantronix® Xport™ Ethernet to RS-232 converter, which provides the 
4033A’s Ethernet connection. 

The 4033A ships from the factory with a default IP address of 0.0.0.0, which enables DHCP. If 
the network has a DHCP server, it will automatically assign the unit an IP address, gateway 
address, and subnet mask when the unit starts up.  

To monitor multiple 4033As remotely through their Ethernet connections without DHCP re-
assigning IP addresses, you must assign each unit a static IP address. You identify which unit is 
the source of an alarm by its IP address. Each unit has its own unique hardware address, labeled 
MAC address, on a label on the rear panel. Follow the instructions in this section to manually 
assign a unit’s static IP address. 

Note

 

For more detailed information, see the 

Xport™ User Manual. 

Section 3.3 discusses several 

different ways that you can assign IP addresses. Chapter 4 explains how to configure a static IP 
address. You can download the 

Xport™ User Manual 

from the Lantronix® Web site as an 

Adobe® Acrobat® PDF file. Go to: 
http://www.lantronix.com/products/eds/xport/XPort_UM_900-270.pdf 

 

If you want to configure a static IP address of many different 4033As, you can install the 
Lantronix® DeviceInstaller software. This software is available only by downloading from the 
Lantronix® Web site. Go to: 

http://www.lantronix.com/support/utils/dst/index.html

. Alternatively 

Содержание TSC 4033A

Страница 1: ...Operations and Maintenance Manual 1 TSC 4033A 1 x 15 5 Volt Pulse Distribution Amplifier Operations and Maintenance Manual 4775 Walnut Street Suite 1B Boulder CO 80301 USA www timing com Phone 303 939...

Страница 2: ...ier Operations and Maintenance Manual Copyright 2004 Timing Solutions Corporation Timing Solutions is a trademark of the Timing Solutions Corporation Other product and company names may be trademarks...

Страница 3: ...e 4033A 14 3 1 Accessing the System 14 3 2 Checking System Status 14 3 2 1 Checking Alarms and Input Frequency 14 3 2 2 Checking Model Number and Software Version 15 3 3 Understanding Alarm Output 15...

Страница 4: ...This page intentionally left blank...

Страница 5: ...bols appear throughout the manual as well as on the unit itself Note This symbol means the following information is a note that gives you important information that may affect how you use the 4033A Ca...

Страница 6: ...TWORK DO NOT CONNECT TO TELECOM CONNECTIONS THAT CARRY HAZARDOUS VOLTAGES 101010 Serial port connection DO NOT CONNECT TO TELECOM CONNECTIONS THAT CARRY HAZARDOUS VOLTAGES Earth terminal symbol Used t...

Страница 7: ...ow to replace fuses and verify operational problems Chapter 5 Warranty and Shipping Information on page 17 explains how to contact Timing Solutions Corporation for warranty service and provides shippi...

Страница 8: ...green LED status for the input and green red for all output signals as well as indicating an input of 1 PPS DC IRIG or Other A green LED on the front panel also provides power status Figure 1 shows th...

Страница 9: ...rify that input line voltage and current capacity are within specifications before turning on power to the unit Disconnect all sources of input power before removing the top cover of this unit Operati...

Страница 10: ...ounting screws Screwdriver for the rack mount screws and slide as needed 2 4 3 Making Connections 2 4 3 1 Input Power AC INPUT POWER The input power to the unit is supplied through a detachable 3 pron...

Страница 11: ...ult IP address of 0 0 0 0 which enables DHCP If the network has a DHCP server it will automatically assign the unit an IP address gateway address and subnet mask when the unit starts up To monitor mul...

Страница 12: ...ress dialog box 5 Type the new IP address for the appropriate Hardware or Ethernet address and click Set IP address The new IP address appears in the Lantronix DeviceInstaller window 6 Test the IP add...

Страница 13: ...ce Manual 13 Note If you move the 4033A to a different network hub after setting up the static IP address the host computer may not be able to make a connection You may need to release the IP address...

Страница 14: ...e of input signal To check system status Type S case sensitive You do not need to type a carriage return or line feed The system returns Scccc llll x r n where cccc is current alarms llll is latched a...

Страница 15: ...stem returns Vxx r n where xx is the software version Example V80 Sending an I command will turn on the three front panel input LEDs for 2 seconds before returning to normal operation This provides th...

Страница 16: ...cement fuse for a standard IEC 320 power entry module with fuse 5 x 20 mm 1 A 250 V Time lag fuse To replace a fuse 1 Disconnect the power cable from the back of the 4033A 2 Using a small screwdriver...

Страница 17: ...4033A Operations and Maintenance Manual 17 This page intentionally left blank...

Страница 18: ...Corporation 4775 Walnut Street Suite 1B Boulder CO 80301 USA 5 2 Shipping Information If you need to ship this system for any reason including returning equipment to Timing Solutions for warranty serv...

Страница 19: ...4033A Operations and Maintenance Manual 19 This page intentionally left blank...

Страница 20: ...uipment for Measurement Control and Laboratory Use Part 1 General Requirements EMC EN61326 2000 Electrical Requirements for Electrical Equipment for Measurement Control and Laboratory Use Part 1 Gener...

Страница 21: ...4033A Operations and Maintenance Manual 21 This page intentionally left blank...

Страница 22: ...eed 10 of nominal supply voltage Power Inlet Type IEC 60320 sheet C14 Power Supply Cord Set 18 AWG 0 75 mm2 minimum Power Mains Fuse 2 250V 1A Time lag 5x20 mm Signal Input Impedance 50 Logic one 2 4...

Страница 23: ...ment suitable for continuous operation Table 4 lists the environmental specifications for the 4033A Table 4 Environment specifications Item Temperature Relative Humidity Altitude In Use 0 C to 50 C 0...

Страница 24: ...P Dynamic Host Configuration Protocol ESD electrostatic discharge LED light emitting diode LSB least significant bit MSB most significant bit PDF portable document format PWA printed wiring assembly R...

Страница 25: ...checking 13 input power connecting 10 input signal connecting 10 installing a 4033A 10 instructions packing 17 safety 9 IP address assigning 11 L LAN connecting 11 LAN port symbol 6 Lantronix DeviceIn...

Страница 26: ......

Страница 27: ...4033A Operations and Maintenance Manual 1...

Отзывы: