Time Electronics 1021 Скачать руководство пользователя страница 8

1021 Technical Manual 

4. 

E lectrical Nois e P ick-Up 

In addition to the inherent electrical noise from internal circuitry of the instrument, the output 
can contain fluctuations with originate from outside sources.  The effects of noise on current 
sources in % or parts per million (P.P.M.) in general, this fraction becomes larger as the 
signal level is reduced. 
 
The 1021 current source has a very high output resistance (greater than 100M ohms).  As 
will all high impedance circuits it is prone to pick-up of electrical signals via capacitance 
coupling.  Voltage sources do not suffer from this problem since they have, in general, very 
low output resistances (less than 1 ohm). 
 
The output terminals of the 1021 can be compared with the input of a high input impedance 
audio amplifier as far as noise pick-up is concerned.  Therefore, for high performance work 
at low current levels it is important to use screened leads.  It is also important to consider 
the effects of earthing when using the 1021 to calibrate mains powered equipment. 
 
The major source of electrical noise pick-up is the 50/60Hz mains.  This noise is generally 
present to a greater or lesser extend in all environments where mains power is available.  
This noise pick-up is capacitively coupled and is AC.  In most applications, although 
superimposed the DC output, it will not affect the DC accuracy.  This is the case with small 
amount of pick-up; but at high levels (several volts peak to peak) the output circuitry of the 
1021 is overloaded by the noise signal and it converts it into a DC offset with results in an 
error in the output. 
 
It is important to realise that when measuring electrical noise at the output terminals of the 
1021, the voltage noise measurements must be divided by the load resistance to give the 
current noise. 
 
A certain amount of noise is generated internally by the 1021 and when high load 
resistances are used this can be seen on the output terminals as a series of voltage spikes.  
The effect of this voltage noise on the output current is very small and does not affect the 
DC accuracy. 
Two of the more common sources of noise are discussed below: 

 
Mains Power Lines 

As has been stated previously, mains power lines are probably the largest source of noise 
often covering a wide spectrum of frequencies.  The most effective solution is to locate its 
source and provide suitable filtering such as capacitor suppression on arcing contacts. 
This noise is usually produced in one of the following ways: 

1. 

50/60Hz signal pick-up

  -  This is generally caused by the close proximity of unscreened 

mains cables or mains powered equipment.  The effects of this type of noise can usually be 
considerably reduced by correct screening and earthing. 
2. 

Transient signal pick-up

 -  This is generally caused by heavy load switching on the 

mains such as electric motors, electric ovens, etc.  Its effect is to cause a transient variation 
in the output which can last longer than the actual duration of the noise.  This is because the 
instrument has been transiently overloaded and requires time to recover. 

 
Common Noise Mode

 

Additional noise and variation of the output current can be caused by large, common mode 
voltages.  These occur when the instrument is used to calibrate any input which is above 
ground potential or has an AC component with respect to ground.  Under normal 
circumstances, it is recommended that a 100v DC common mode is not exceeded.  The AC 
common mode rejection is determined by the capacitive unbalance to ground of the output 
terminals and associated connections. 
  

Содержание 1021

Страница 1: ...otany Industrial Estate Tonbridge Kent TN9 1RH Tel 44 0 1732 355993 Fax 44 0 1732 770312 Email mail TimeElectronics co uk Web Site www TimeElectronics com Time Electronics 1021 Current Source With Null Measuring Facility Technical Manual ...

Страница 2: ...echarging 7 4 E lectrical Nois e Pick Up 8 5 Circuit Des cription 9 5 1 Power Supply 9 5 2 Mains Recharger 9 5 3 Circuit Module and Range Switch 9 6 Maintenance and R epair 10 6 1 Maintenance 10 6 2 Repair 10 7 R ecalibration 11 7 1 Calibration Procedure 12 7 2 Linearity 13 8 Guarantee S ervicing 14 All Time Electronics instruments are subject to continuous development and improvement and in conse...

Страница 3: ...en used even if a contact fails the 1021 will still operate correctly The accuracy and stability are such that a wide range of applications are possible In the process industries it may be used to test and calibrate current sensitive transducers and their associated indicating and recording instruments The semiconductor industry requires constant current sources for parameter measurements It may a...

Страница 4: ...m per ºC 10 ºC to 50 ºC Better than 25 ppm per hr at constant temperature Null Sensitivity Adjustable from 25mA to 25uA Load Regulation Better than 20 ppm per volt change in output Power Supply 8 NiMH rechargeable batteries with external mains re charger Standard mains voltage is 220 250V 50 60Hz 100 125V 50 60Hz is available but must be specified on order Dimensions 200 x 75 x 110mm 215 x 100 x 1...

Страница 5: ...5 1021 Technical Manual 5 2 Controls 1 8 9 7 3 5 2 6 4 10 11 12 ...

Страница 6: ...n each decade 8 Source Null Switch A two position toggle switch for selecting the mode of operation In the Source position the instrument operates as a current source In the Null position the instrument is used for the measurement of current 9 Supply On Off Switch and Range Switches Four press button switches with interlocking on the three range switches to ensure that only one range can be select...

Страница 7: ...the unknown current equals the digit switch setting 3 4 Voltage Compliance Adjustment V O P The minimum voltage compliance of approximately 14v is obtained by turning the V O P control fully anti clockwise With full clockwise adjustment approximately 40v is obtained The voltage level is checked by measuring across the output terminals when the voltage limit LED is illuminated It is recommended tha...

Страница 8: ... terminals of the 1021 the voltage noise measurements must be divided by the load resistance to give the current noise A certain amount of noise is generated internally by the 1021 and when high load resistances are used this can be seen on the output terminals as a series of voltage spikes The effect of this voltage noise on the output current is very small and does not affect the DC accuracy Two...

Страница 9: ...ife of a fully charged battery should be up to 20 hours 5 2 Mains Recharger The mains recharger is a separate unit It will fully charge the batteries in 12 to 16 hours The standard recharger requires a mains input of 200V to 240V at 50 60Hz A recharger requiring 100V to 120V at 50 60Hz is also available but should be specified when ordering The recharger is connected to the 1021 by a miniature 2 p...

Страница 10: ...ensions 20mm length by 5mm diameter The fuses are available from Time Electronics Ltd or your local supplier 6 2 Repair NOTE No repair work should be undertaken by the customer while the instrument is under warranty as such work may render the warranty invalid Some of the precision components used in this instrument are not readily available and make repairs by the customer difficult if these comp...

Страница 11: ... available before attempting recalibration There are 2 ways the 1021 can be recalibrated a The output current can be converted to a voltage by connecting a precision resistor across the output terminals 3 values are required 100 1K and 10K ohms The resistors must be accurate to better than 0 01 and have a good stability under load The 10 ohm will dissipate 0 1 watts at 100mA and therefore must hav...

Страница 12: ...ll digits to zero adjust the ZERO trimmer to give 1 micro amp output For recalibration method a the precision 1K ohm resistor should be connected to the output terminals Note the zero adjustment is unipolar i e it is not possible to adjust the output to less than zero 2 Select the 1mA range and set all digits to 9999 Adjust the 1MACAL trimmer to bring the output to exactly 1mA For recalibration me...

Страница 13: ...ng conditions the high performance of these resistors ensures that the 1021 linearity will remain within specification for the instrument s lifetime Should the resistors be damaged or become faulty a replacement procedure should be adopted ensuring that the replacement resistor s is matched to the others on that digit bank The 1K ohm values should be matched to 0 01 the 100 ohms to 0 05 the 10 ohm...

Страница 14: ...pecialised knowledge needed to keep your instrument in peak condition and we also maintain a comprehensive spare parts service Please enclose details of the service required and your full company details including a contact name when returning for servicing Returning Instruments When returning instruments please ensure that they have been adequately packed preferably in the original packing suppli...

Отзывы: