1021 Technical Manual 

3. 

Operation 

3.1.  Preliminaries 

The 1021 is supplied complete with rechargeable batteries (housed in the instrument) and a 
mains charger. Before operation, the batteries should be checked on the battery level 
indicator.  It is important to recheck the battery condition when the instrument is supplying 
currents in excess of a few tens of mA, since the additional load may reduce the battery 
voltage to below the minimum level. 
  

3.2.  Normal Operation 

Using the specification together with the front panel controls makes the operation self-
explanatory.  It is important to understand the voltage capability and the voltage limit 
indicator of the instrument to ensure that requirements are within limits. 
 

3.3.  Measurement of Current 

1. Select NULL on the front panel function switch. 
2. Select NORM output. 
3. Adjust SENS fully anti-clockwise (minimum sensitivity). 
4. Select RANGE consistent with the current to be measured. 
5. Connect the unknown current such that it flows into the + ve terminal. 
6. Adjust the digit switch and sensitivity control for null balance on the LED indicator  

(green centre LED). 

7. When balance at the required sensitivity is reached, the value of the unknown current   

equals the digit switch setting. 

  

 

3.4.  Voltage Compliance Adjustment    -   V O/P 

The minimum voltage compliance of approximately 14v is obtained by turning the V O/P 
control fully anti-clockwise.  With full clockwise adjustment approximately 40v is obtained. 
  
The voltage level is checked by measuring across the output terminals when the voltage 
limit LED is illuminated. 
It is recommended that the voltage level is set to the minimum required for the work in hand. 
  

3.5.  Overload Conditions 

The instrument can withstand either a short or open circuit on the output terminals 
indefinitely.  Overload conditions can result if an attempt is made to drive current into the 
instrument by applying a voltage to the output terminals.  In most cases this will cause the 
output fuse to blow but could still damage the output circuitry. 
  

3.6.  Battery Recharging 

Under normal conditions the 1021 should be switched off during battery recharge.  Full 
recharge takes 10-16 hours.  The battery charging rate is such that the battery cannot be 
damaged by overcharging. 
  
It is not recommended that the 1021 be operated during battery recharge although output 
currents of a few milliamps at low compliance voltage can be obtained without discharging 
the batteries. It should be noted however that additional electrical noise will appear on the 
output due to the indirect mains power connection. 
 

 

Содержание 1021

Страница 1: ...otany Industrial Estate Tonbridge Kent TN9 1RH Tel 44 0 1732 355993 Fax 44 0 1732 770312 Email mail TimeElectronics co uk Web Site www TimeElectronics com Time Electronics 1021 Current Source With Null Measuring Facility Technical Manual ...

Страница 2: ...echarging 7 4 E lectrical Nois e Pick Up 8 5 Circuit Des cription 9 5 1 Power Supply 9 5 2 Mains Recharger 9 5 3 Circuit Module and Range Switch 9 6 Maintenance and R epair 10 6 1 Maintenance 10 6 2 Repair 10 7 R ecalibration 11 7 1 Calibration Procedure 12 7 2 Linearity 13 8 Guarantee S ervicing 14 All Time Electronics instruments are subject to continuous development and improvement and in conse...

Страница 3: ...en used even if a contact fails the 1021 will still operate correctly The accuracy and stability are such that a wide range of applications are possible In the process industries it may be used to test and calibrate current sensitive transducers and their associated indicating and recording instruments The semiconductor industry requires constant current sources for parameter measurements It may a...

Страница 4: ...m per ºC 10 ºC to 50 ºC Better than 25 ppm per hr at constant temperature Null Sensitivity Adjustable from 25mA to 25uA Load Regulation Better than 20 ppm per volt change in output Power Supply 8 NiMH rechargeable batteries with external mains re charger Standard mains voltage is 220 250V 50 60Hz 100 125V 50 60Hz is available but must be specified on order Dimensions 200 x 75 x 110mm 215 x 100 x 1...

Страница 5: ...5 1021 Technical Manual 5 2 Controls 1 8 9 7 3 5 2 6 4 10 11 12 ...

Страница 6: ...n each decade 8 Source Null Switch A two position toggle switch for selecting the mode of operation In the Source position the instrument operates as a current source In the Null position the instrument is used for the measurement of current 9 Supply On Off Switch and Range Switches Four press button switches with interlocking on the three range switches to ensure that only one range can be select...

Страница 7: ...the unknown current equals the digit switch setting 3 4 Voltage Compliance Adjustment V O P The minimum voltage compliance of approximately 14v is obtained by turning the V O P control fully anti clockwise With full clockwise adjustment approximately 40v is obtained The voltage level is checked by measuring across the output terminals when the voltage limit LED is illuminated It is recommended tha...

Страница 8: ... terminals of the 1021 the voltage noise measurements must be divided by the load resistance to give the current noise A certain amount of noise is generated internally by the 1021 and when high load resistances are used this can be seen on the output terminals as a series of voltage spikes The effect of this voltage noise on the output current is very small and does not affect the DC accuracy Two...

Страница 9: ...ife of a fully charged battery should be up to 20 hours 5 2 Mains Recharger The mains recharger is a separate unit It will fully charge the batteries in 12 to 16 hours The standard recharger requires a mains input of 200V to 240V at 50 60Hz A recharger requiring 100V to 120V at 50 60Hz is also available but should be specified when ordering The recharger is connected to the 1021 by a miniature 2 p...

Страница 10: ...ensions 20mm length by 5mm diameter The fuses are available from Time Electronics Ltd or your local supplier 6 2 Repair NOTE No repair work should be undertaken by the customer while the instrument is under warranty as such work may render the warranty invalid Some of the precision components used in this instrument are not readily available and make repairs by the customer difficult if these comp...

Страница 11: ... available before attempting recalibration There are 2 ways the 1021 can be recalibrated a The output current can be converted to a voltage by connecting a precision resistor across the output terminals 3 values are required 100 1K and 10K ohms The resistors must be accurate to better than 0 01 and have a good stability under load The 10 ohm will dissipate 0 1 watts at 100mA and therefore must hav...

Страница 12: ...ll digits to zero adjust the ZERO trimmer to give 1 micro amp output For recalibration method a the precision 1K ohm resistor should be connected to the output terminals Note the zero adjustment is unipolar i e it is not possible to adjust the output to less than zero 2 Select the 1mA range and set all digits to 9999 Adjust the 1MACAL trimmer to bring the output to exactly 1mA For recalibration me...

Страница 13: ...ng conditions the high performance of these resistors ensures that the 1021 linearity will remain within specification for the instrument s lifetime Should the resistors be damaged or become faulty a replacement procedure should be adopted ensuring that the replacement resistor s is matched to the others on that digit bank The 1K ohm values should be matched to 0 01 the 100 ohms to 0 05 the 10 ohm...

Страница 14: ...pecialised knowledge needed to keep your instrument in peak condition and we also maintain a comprehensive spare parts service Please enclose details of the service required and your full company details including a contact name when returning for servicing Returning Instruments When returning instruments please ensure that they have been adequately packed preferably in the original packing suppli...

Отзывы: