Maintenance and Service
Danger of possibly contaminated parts!
Contaminated parts can cause personal injuries. Inform your-
self regarding possible contamination before you start working.
Be sure to follow the relevant instructions and take care of nec-
essary protective measures.
The unit requires no maintenance. External dirt and soiling can be removed by a
damp cloth.
Should a defect or damage occur on your VSP/VCP, please send the instrument
to us for repair and enclose a decontamination declaration.
The unit is not prepared for customer repair!
Defective sensor heads can be exchanged on-site by calibrated
replacement sensors (spare parts B_VSP63D, B_VSP64D,
B_VCP63D).
Malfunction of the unit, which is caused by contamination or
wear and tear is not covered by warranty.
Error messages and malfunction
Problem
Possible Cause
Correction
high measurement error
contamination, ageing,
extreme temperature,
maladjustment
readjustment,
replace sensor or
send unit for repair
Pirani zero adjustment not
possible
measurement error ex-
ceeds possible range of
readjustment
replace sensor or
send unit for repair
"UR" via RS485
pressure under range
(pressure < 1x10
-4
mbar)
"ERROR1" via RS485
Status LED continuously red
defective electronics or
sensor
send unit for repair or
replace sensor
Access code "7" via RS485 Error message / warning
see separate documenta-
tion Thyracont Communica-
tion Protocol
Содержание VCP63E
Страница 1: ...VSP63E VSP64E VCP63E Vakuum Transmitter Vacuum Transducer Betriebsanleitung Operating Instructions ...
Страница 2: ...2 VSPE de 171117 ...
Страница 19: ...19 VSPE de 171117 ...
Страница 22: ...22 VSPE de 171117 Konformitätserklärung ...
Страница 39: ...39 VSPE de 171117 ...
Страница 42: ...42 VSPE de 171117 Declaration of Conformity ...
Страница 43: ...43 VSPE de 171117 ...
Страница 44: ...VSPE de 171117 ...