background image

Premier

 Ankle Brace

4615 Shepard Street Bakersfield, CA 93313  

Tele 800.432.3466 - Fax 800.798.2722

www.ThuasneUSA.com

MADE IN USA BY

TOWNSEND DESIGN

4615 SHEPARD STREET

BAKERSFIELD, CA 93313 

TEL 661.837.1795

Medical Professional

 Instruction Guide 

Joint

 Adjustments  

(To Fixate The Joints Or Limit Inversion/Eversion Motion)

For braces ordered with standard joints, the adjustment 

process involves removing a screw, joint cover plate and the 

factory installed insert. A stop tree is included with each brace. 

The fitter can select the optimum inserts for adjusting the 

inversion/eversion motion of the carbon graphite ankle joints. 

After the appropriate insert is installed, put the cover plate 

back on and lock it into position with the screw.

Inversion/Eversion Adjustment Kit

-5 Degrees

-5

o

+5

o

-2.5 Degrees

-2.5

o

+2.5

o

Free Motion

Free 

Motion

+5 Degrees
+2.5 Degrees
0 (Fixed)

0

o

0

o

The joints provide unlimited dorsiflexion and plantarflexion 

motion. Inversion and/or eversion can be limited or fixed by 

installing Delrin inserts to one or both joints. 

Fitting

 Instructions

• 

Place the brace inside the 

shoe before putting the 

brace on the patient’s foot.

Posterior Tibia Shell

• 

Slip the patient’s foot into the 

brace (inside the shoe) so the 

heel fits comfortably into the 

molded area on the footplate.

• 

Secure the uprights to the 

leg using the two straps. 

Trim Velcro straps to make 

it easier for the patient to 

don and doff the brace.

• 

After the initial fitting, ensure 

the patient feels comfortable 

in the brace, with no sharp 

pressure points. If they 

feel the brace rubbing their 

skin at any time, subtle 

modifications can be made to 

the footplate (if you ordered 

a plastic foot) with a heat gun 

to relieve pressure points.

• 

Instruct the patient to initially 

wear the brace 1-2 hours 

the first few days. Then 

gradually increase the time 

each day. Overuse can cause 

aching and discomfort.

Anterior Tibia Shell

• 

Step through the tibia shell 

and the molded footplate 

so the heel rests firmly and 

intimately against the heel of 

the footplate and tibia shell. 

• 

Secure the upright to the leg 

using the one posterior strap. 

The patient will need to loosen 

and tighten this strap each time 

they put on or take off the brace. 

Trim the length is it is too long.

Warranty

 Information

30-day free remold/remake service (from the date of fabrication) 

to resolve any initial fitting issues. A new cast or leg scan may be 

required. For one year from the date of sale, if the graphite shell, 

foot plate or hinges fail or are structurally compromised during 

routine and customary activities, Townsend will routinely repair 

or replace the brace at no charge. Replacement parts including 

straps and pads are guaranteed against defects and unnatural 

wear for six months.

L-0085 REV. A

Отзывы: