MyoTrac Infiniti
™
User Guide
18
Attention: When you insert the extender cable (lead wire) into the electrode
connector, MAKE SURE THAT NO BARE METAL OF THE PINS IS EXPOSED.
Attention:
When using stimulation ensure that the intensity is set below the maximum for the
selected electrode size. This is assuming an electrode impedance of 500
or 1k
respectively.
1’’ Round = Max 69mA / 49mA
2’’ x 2’’ Square = Max 100mA / 100mA
3’’ x 4’’ Rectangle = Max 100mA / 100mA
St Cloud Vaginal = Max 100mA / 74mA
Femelex = Max 100mA
Note:
If an external pre-amplifier is connected to the device all stim settings are disabled to protect
the circuitry of the amplifier.