Operating Instruction
Setup the Wavefront Sensor to work using the internal reference. Do the appropriate
setting within the
.
Adjust the point source straight in front of the Wavefront Sensor at a distance in the
range 10 cm to 5 m. The Wavefront Sensor will measure the distance and displays it
as Radius of Curvature RoC.
Larger distances are better suited for calibration because the wavefront curvature is
lower and closer to a plane wavefront
Start the calibration procedure using
Menu
Calibration
Spherical
Wavefront
or simply click on the appropriate symbol
in the toolbar. The following
panel illustrates the required calibration setup.
Click '
Do Calibration
' to perform the user calibration.
In case you are currently working with an user reference you will reminded to switch
to the internal reference.
65
Содержание WFS Series
Страница 2: ...2012 Operation Manual Optical Wavefront Sensors Shack Hartmann Sensors WFS Series ...
Страница 3: ...Version Date 3 5 March27 2012 2007 2012 Thorlabs GmbH 2007 2012 Thorlabs GmbH ...
Страница 72: ... 2007 2012 Thorlabs GmbH 69 Operating Instruction ...
Страница 132: ... 2007 2012 Thorlabs GmbH 129 Appendix TypicalResponse of the WFS CCD camera TypicalResponse of the WFS10 CMOS camera ...
Страница 139: ... 2007 2012 Thorlabs GmbH 136 WFS Series WFS10 Adapter Plate ...
Страница 143: ... 2007 2012 Thorlabs GmbH 140 WFS Series 6 3 9 Drawing WFS150 5C ...
Страница 144: ... 2007 2012 Thorlabs GmbH 141 Appendix 6 3 10 Drawing WFS150 7AR ...
Страница 145: ... 2007 2012 Thorlabs GmbH 142 WFS Series 6 3 11 Drawing WFS300 14AR ...
Страница 146: ... 2007 2012 Thorlabs GmbH 143 Appendix 6 3 12 Drawing WFS10 5C ...
Страница 147: ... 2007 2012 Thorlabs GmbH 144 WFS Series 6 3 13 Drawing WFS10 7AR ...
Страница 148: ... 2007 2012 Thorlabs GmbH 145 Appendix 6 3 14 Drawing WFS10 14AR ...
Страница 156: ......