© 2016 Thorlabs
3 External Connections
13
·
Pin 10 and / or Pin 11 are not connected:
The resistance measurement is done as 2-wire measurement, the resistances of cable and
connectors are included in the measurement. The temperature control works at normal
conditions.
·
Pin 3 and / or Pin 4 are not connected:
The resistance measurement is done as 2-wire measurement, the resistances of cable and
connectors are included in the measurement. The temperature control works at normal
conditions.
3.8 Connecting a TEC Element with Voltage Detector
The TEC element is connected between pins 5, 6, 7 (plus pole) and pins 13, 14, 15 (minus pole).
For a 4-point measurement of the TEC voltage, connect the TEC element to pin 2 and 9 to
measure the voltage directly at the TEC element.
Pin 2 and pin 9 may also be connected directly to the plug at the temperature module (i.e. pin 2
with pin 5,6,7 and Pin 9 with pin 13,14,15), but this may lead to a measurement error due to a
voltage drop across the cable to the TEC element, caused by the TEC current. The indicated
voltage will then be slightly higher.
4-pole Measurement of the TEC Voltage
Attention
An reverse poled TEC element may lead to a thermal runaway and destruction of the connected
components. Please refer to section
Polarity Check of the TEC Element
.
14
Содержание TED8000 Series
Страница 1: ...PRO800 PRO8000 4 Series Modular Temperature Controllers TED8000 Operation Manual 2016 ...
Страница 2: ...Version Date Item No 3 2 08 Sep 2016 M0009 510 227 Copyright 2016 Thorlabs ...
Страница 42: ... 2016 Thorlabs 40 TED8000 Structure of the registers DEC DEE and EDE ...
Страница 43: ... 2016 Thorlabs 6 Communication with a PC 41 Structure of the registers ESR ESE STB and SRE ...
Страница 59: ......
Страница 60: ...www thorlabs com ...