6.3.1 Select the Module Slot
Command
Explanation
Response Example
":SLOT <NR1>"
Selects a slot for further programming
<NR1>
= 1…8 PRO8000, PRO8000-4
<NR1>
= 1…2 PRO800
":SLOT?"
Queries the selected slot
[:SLOT <NR1><LF>]
6.3.2 Thermistor Calibration - Exponential Method
Command
Explanation
Response Example
Programming
":CALTB:SET <NR3>"
Programs the energy constant B
val
":CALTR:SET <NR3>"
Programs the nominal resistance R
0
":CALTT:SET <NR3>"
Programs the nominal temperature T
0
Reading
":CALTB:SET?"
Reads the energy constant B
val
[:CALTB:SET <NR3><LF>]
":CALTR:SET?"
Reads the nominal resistance R
0
[:CALTR:SET <NR3><LF>]
":CALTT:SET?"
Reads the nominal temperature T
0
[:CALTT:SET <NR3><LF>]
":CALTB:MIN?"
Reads the minimum allowed energy constant B
val
[:CALTB:MIN <NR3><LF>]
":CALTR:MIN?"
Reads the minimum allowed nominal resistance R
0
[:CALTR:MIN <NR3><LF>]
":CALTT:MIN?"
Reads the minimum allowed nominal temperature T
0
[:CALTT:MIN <NR3><LF>]
":CALTB:MAX?"
Reads the maximum energy constant B
val
[:CALTB:MAX <NR3><LF>]
":CALTR:MAX?"
Reads the maximum allowed nominal resistance R
0
[:CALTR:MAX <NR3><LF>]
":CALTT:MAX?"
Reads the maximum allowed nominal temperature T
0
[:CALTT:MAX <NR3><LF>]
Refer to section
Thermistor Calibration - Exponential Method
.
Attention
These commands
do not
apply to TED8000-KRYO!
29
Содержание TED8000 Series
Страница 1: ...PRO800 PRO8000 4 Series Modular Temperature Controllers TED8000 Operation Manual 2016 ...
Страница 2: ...Version Date Item No 3 2 08 Sep 2016 M0009 510 227 Copyright 2016 Thorlabs ...
Страница 42: ... 2016 Thorlabs 40 TED8000 Structure of the registers DEC DEE and EDE ...
Страница 43: ... 2016 Thorlabs 6 Communication with a PC 41 Structure of the registers ESR ESE STB and SRE ...
Страница 59: ......
Страница 60: ...www thorlabs com ...