Chapter 4: Setup and Operation
Page 14
CTN008800-D02
4.11. Typical Performance
The time it takes to open the shutter or close the shutter, IA and DE as shown in
Figure 7 on page 13, are constant characteristics of the device. The time delay for
both opening and closing is a result of the time it takes to energize the coils. The
net effect is the delay from the initiation of the starting pulse to the shutter
responding (IA) is typically 11 ms. Similarly, the closing time (DE) is typically 15
ms.
The mechanical rise time of the shutter is given as AC in Figure 7 on page 12. The
fall time is given as EG. The typical values for the rise and fall times respectively
are 12 ms and 9 ms.
Therefore, the total time it typically takes for the shutter to open would be IC or
21 ms. The total time it typically takes for the shutter to close would be DG or
24 ms. Depending on your application, some or all of these values are important
to note.