5.6. Troubleshooting 

As shown in Figure 10, there are 2 LEDs on the front panel of the controller 

  C3. Bulb status indication LED 

  C4. Temperature warning LED 

During normal operation, all LEDs will turn GREEN. 

5.6.1. Overheating 

On occasions when the controller is overheated, the temperature warning LED, C4, will turn red and the lamp will 
be shut off automatically. 

The LCD screen of the controller will give an error message of “OT” (over temperature). 

When such a warning occurs, shut down the controller and allow it some time to cool down before switching it on 
again. 

If the warning keeps occurring when the controller is operating within the specified temperature range, please 
contact Thorlabs Tech Support. 

5.6.2. Lamp Not 

C

onnected 

If the controller is switched on without being connected to the lamp, the bulb status indication LED, C3, will turn 
RED and the LCD screen will give an error message of “Power Off, Check Lamp”. Please switch off the controller 
and recheck the connection between the lamp housing and controller. 

If the warning message continues to occur even when the lamp housing is properly connected, please contact 
Thorlabs Tech Support. 

5.6.3.  Bulb State Warning 

On occasions when the bulb can not be ignited, the bulb status indicating LED, C3, will turn RED. And the LCD 
screen will give an error message of “Fault”. 

A bulb state warning is usually caused by the following reasons: 

1.  No bulb is installed. 

2.  The bulb is installed incorrectly. 

3.  The bulb has reached the end of its lifetime. 

4.  Failed attempt to ignite a hot bulb. 

When such a warning occurs, please double-check that the bulb is securely seated in the holder as well as the 
lifetime of the bulb. Also make sure the lamp is cooled down before attempting to switch the unit on. 

If the warning keeps occurring with a new bulb even when it is cold, please contact Thorlabs Tech Support. 

 

Содержание SLS401

Страница 1: ...SLS401 SLS402 Benchtop Arc Lamp User Guide...

Страница 2: ...anel 8 4 3 3 Controller Back Panel 9 4 4 Bulb Handling 9 4 5 Bulb Installation 10 4 6 Setting Up on a Work Surface 13 Chapter 5 Operation 16 5 1 Starting the Lamp 16 5 2 Shutter Operation 16 5 3 Adjus...

Страница 3: ...Description Direct Current Alternating Current Both Direct and Alternating Current Earth Ground Terminal Protective Conductor Terminal Frame or Chassis Terminal Equipotentiality On Supply Off Supply I...

Страница 4: ...lectric shock resulting in severe injury or even death Do not operate without cover installed WARNING This unit must not be operated in explosive environments WARNING Do not operate in wet damp condit...

Страница 5: ...re T K is the Boltzmann constant h is the Planck constant c is the speed of light By measuring the spectrum of the black body radiation we can calculate the temperature of this black body Figure 1 sho...

Страница 6: ...cury Xenon Arc Lamps Both bulbs are filled with Xenon gas while the Mercury Xenon bulb has traces of mercury mixed in The Xenon arc lamp provides a continuous emission over the 240 nm to 2400 nm spect...

Страница 7: ...the arc lamp bulb as shown in Figure 4 This configuration focuses the light from arc back back to the arc position so that it can be be collimated by the front lens Figure 4 gives an outline of the o...

Страница 8: ...ss tc 8 mm Edge Thickness te 2 87 mm Material UV Grade Fused Silica Design Wavelength 587 6 nm Figure 6 Collimating Lens Specifications The silver concave mirror is oriented with the coated concave su...

Страница 9: ...or Connecting Lamp Housing and Controller 1 Power Cord 1 4 mm Hex Key 1 1 3 mm Hex Key 4 2 Unpacking The packaging of the SLS401 and SLS402 is illustrated below Figure 8 Packaging of the SLS401 and SL...

Страница 10: ...Housing L1 Shutter Control Knob L2 Collimating Lens L3 4 40 Threaded Hole for 60 mm Cage Systems L4 Adjustable Feet L5 Knobs for Adjusting Rear Concave Mirror Alignment 4 3 2 Controller Front Panel Fi...

Страница 11: ...ng handling When handled correctly the bulb is a very safe instrument However improper handling may cause hazard to users The bulb of an arc lamp contains high internal pressure and would burst if han...

Страница 12: ...temperature may reach a few hundred C during operation Always wait at least 30 minutes after turning off the lamp for cool down before performing a bulb replacement WARNING Always wear protective glo...

Страница 13: ...quipped with a dummy bulb to secure the top bulb holder during transportation Loosen the thumb screws on top and bottom bulb holder by turning counterclockwise and remove the dummy bulb Keep the dummy...

Страница 14: ...nsert the bulb with anode pointing up Install the upper bulb holder on the anode Make sure the anode is securely inserted into the holder and then tighten both thumb screws by turning clockwise Figure...

Страница 15: ...ne is included with the SLS401 and SLS402 to loosen and remove the cap screws Figure 17 Removing Caps from the Adjustable Feet Use a 10 mm wrench not included to rotate the washer underneath the botto...

Страница 16: ...to an optical breadboard or table It is compatible with both metric and imperial breadboards tables Simply remove all four feet with a 4 mm hex key and use M6 or 1 4 20 cap screws not included in the...

Страница 17: ...tor onto the socket and turn the threaded cap screw clockwise to lock the connector using the included 1 3 mm hex key To remove the cable press the connector against the socket lightly and turn the ca...

Страница 18: ...bulb The end of lifetime is defined as the time when the output optical power of the bulb reaches half of its initial value The actual life time value varies between individual bulbs The timer provid...

Страница 19: ...g Lens Position 5 4 Adjusting the Rear Concave Mirror The concave mirror is shipped preset in the optimum position If adjustment is required remove the side plate as instructed in Section 4 4 to revea...

Страница 20: ...the front panel of the lamp house to allow the integration of Thorlabs 60 mm Cage System The collimating lens also has internal SM2 threads to allow the integration of Thorlabs SM2 mounts and lens tu...

Страница 21: ...the lamp the bulb status indication LED C3 will turn RED and the LCD screen will give an error message of Power Off Check Lamp Please switch off the controller and recheck the connection between the l...

Страница 22: ...402 lamp housings Before replacing the bulb switch off the lamp and allow at least 30 minutes for the lamp to cool down Remove the side plate as instructed in Section 4 4 Loosen the two thumbscrews on...

Страница 23: ...Typical 0 12 Output Beam Diameter 35 mm Beam Divergencec 3 2 Bulb Lifespan Average d 2000 hours Power Supply 90 264 VAC 47 63 Hz Operating Temperature 0 to 45 C Storage Temperature 15 to 70 C Lamp Ho...

Страница 24: ...Places Output Beam Port w 35 mm Diameter Shutter Knob SM2V05 Lens Tube 4 40 Hole For 60 mm Cage System 4 Places 214 7 mm 8 45 62 2 mm 2 45 62 2 mm 2 45 82 0 mm 3 23 76 7 mm 3 02 217 0 mm 8 54 261 2 m...

Страница 25: ...9 Certifications and Compliances This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this dev...

Страница 26: ...labs products such as Pure OEM products that means assemblies to be built into a unit by the user e g OEM laser driver cards Components Mechanics and optics Left over parts of units disassembled by th...

Страница 27: ...on USA Canada and South America Thorlabs Inc sales thorlabs com techsupport thorlabs com Europe Thorlabs GmbH europe thorlabs com France Thorlabs SAS sales fr thorlabs com Japan Thorlabs Japan Inc sal...

Страница 28: ...www thorlabs com...

Отзывы: